Intercomparison of erythemal broadband radiometers calibrated by seven UV calibration facilities in Europe and the USA
A bi-lateral intercomparison of erythemal broadband radiometers was performed between seven UV calibration facilities. The owners calibrations were compared relative to the characterisation and calibration performed at PMOD/WRC in Davos, Switzerland. The calibration consisted in the determination of...
Saved in:
Published in: | Atmospheric chemistry and physics discussions 2008-02, Vol.8 (1), p.2249-2273 |
---|---|
Main Authors: | , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A bi-lateral intercomparison of erythemal broadband radiometers was performed between seven UV calibration facilities. The owners calibrations were compared relative to the characterisation and calibration performed at PMOD/WRC in Davos, Switzerland. The calibration consisted in the determination of the spectral and angular response of the radiometer, followed by an absolute calibration performed outdoors relative to a spectroradiometer which provided the absolute reference. The characterization of the detectors in the respective laboratories are in good agreement: The determination of the angular responses have deviations below ±4% and the spectral responses agree within ±20%. A "blind" intercomparison of the erythemally weighted irradiances derived by the respective institutes and PMOD/WRC showed consistent measurements to within ±2% for the majority of institutes. One institute showed slightly larger deviation of 10%. The differences found between the different instrument calibrations are all within the combined uncertainty of the calibration. |
---|---|
ISSN: | 1680-7367 1680-7375 |