Are ligament-tensioning devices interchangeable? A study of femoral rotation
Abstract Background During total knee arthroplasty (TKA), femoral rotation can be adjusted either in relation to bony landmarks or by tensioning the ligaments with the knee in 90° of flexion. The primary objective of this study was to compare femoral rotations achieved using various ligament-tension...
Saved in:
Published in: | Orthopaedics & traumatology, surgery & research surgery & research, 2016-06, Vol.102 (4), p.S213-S219 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract Background During total knee arthroplasty (TKA), femoral rotation can be adjusted either in relation to bony landmarks or by tensioning the ligaments with the knee in 90° of flexion. The primary objective of this study was to compare femoral rotations achieved using various ligament-tensioning devices. The secondary objective was to compare these femoral rotations to that indicated by the transepicondylar axis (TEA). Material and methods We performed 13 posterior-stabilised TKA procedures using HiFit (Ceraver® ) on cadaver knees. Before performing the posterior condyle cut, we used an original method to measure the femoral rotation induced by five different ligament-tensioning devices (2 with a ratchet mechanism, 1 with screws, 1 force-sensing device, and 1 with spacer blocks) and the central tibio-femoral distance (CTFD). Results Both ratchet tensioners provided significantly greater mean external rotation values ( P = 0.002), of 4.94° and 4.46°, respectively, compared to the force-sensing and spacer tensioners. Significant differences were found across devices for CTFD, with a mean difference of about 2 mm between the ratchet and screw tensioners versus the force-sensing and spacer tensioners. The mean differences in rotations obtained using the tensioners versus the TEA were close to 0° but with standard deviations greater than 4°. Conclusion Femoral rotation was dependent on the distraction force applied to the joint. Tensioners that did not measure the distraction force were associated with greater distraction force and external rotation values. The TEA criterion did not reliably indicate good ligament balance. Level of evidence Experimental study. |
---|---|
ISSN: | 1877-0568 1877-0568 |
DOI: | 10.1016/j.otsr.2016.02.001 |