Loading…
Comparison of noise-magnitude and noise-texture across two generations of iterative reconstruction algorithms from three manufacturers
To compare the noise-magnitude and noise-texture across two generations of iterative reconstruction (IR) algorithms proposed by three manufacturers according to the dose level. Five computed tomography (CT) systems equipped with two generations of IR algorithms (hybrid/statistical IR [H/SIR] or full...
Saved in:
Published in: | Diagnostic and interventional imaging 2019-07, Vol.100 (7-8), p.401-410 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To compare the noise-magnitude and noise-texture across two generations of iterative reconstruction (IR) algorithms proposed by three manufacturers according to the dose level.
Five computed tomography (CT) systems equipped with two generations of IR algorithms (hybrid/statistical IR [H/SIR] or full/partial model-based IR [MBIR]) were compared. Acquisitions on Catphan 600 phantom were performed at 120kV and three dose levels (3-, 7- and 12-mGy). Raw data were reconstructed using standard “soft tissue” kernel for filtered back projection and one iterative level of two generations of IR algorithms. Contrast to-noise-ratio (CNR) was computed using three regions of interest: two of them placed in the low-density polyethylene (LDPE) and Teflon® inserts and another placed on the solid water. Noise power spectrum (NPS) was computed to assess the noise-magnitude (NPS peak) and noise-texture (NPS spatial frequency).
CNR increased significantly in MBIR compared to H/SIR algorithms for General-Electric (GE) Healthcare (45%±12 [SD]) and Philips Healthcare systems (62%±11 [SD]) (P |
---|---|
ISSN: | 2211-5684 2211-5684 |
DOI: | 10.1016/j.diii.2019.04.006 |