Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT
. The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum (NPS) and RSP accuracy has been investigated, as a preliminary s...
Saved in:
Published in: | Physics in medicine & biology 2023-06, Vol.68 (12), p.124001 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-c445t-4b23aae2b2dcf6e3a11c7efbc3fbc3d9df981dd7fccc2edf4f320ec21507cda73 |
---|---|
cites | cdi_FETCH-LOGICAL-c445t-4b23aae2b2dcf6e3a11c7efbc3fbc3d9df981dd7fccc2edf4f320ec21507cda73 |
container_end_page | |
container_issue | 12 |
container_start_page | 124001 |
container_title | Physics in medicine & biology |
container_volume | 68 |
creator | Fogazzi, Elena Trevisan, Diego Farace, Paolo Righetto, Roberto Rit, Simon Scaringella, Monica Bruzzi, Mara Tommasino, Francesco Civinini, Carlo |
description | . The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum (NPS) and RSP accuracy has been investigated, as a preliminary step to implement a new cross-calibration method for x-ray CT (xCT).
. The INFN pCT apparatus, made of four planes of silicon micro-strip detectors and a YAG:Ce scintillating calorimeter, reconstructs 3D RSP maps by a filtered-back projection algorithm. The imaging performances (i.e. spatial resolution, NPS and RSP accuracy) of the pCT system were assessed on a custom-made phantom, made of plastic materials with different densities ((0.66, 2.18) g cm
). For comparison, the same phantom was acquired with a clinical xCT system.
. The spatial resolution analysis revealed the nonlinearity of the imaging system, showing different imaging responses in air or water phantom background. Applying the Hann filter in the pCT reconstruction, it was possible to investigate the imaging potential of the system. Matching the spatial resolution value of the xCT (0.54 lp mm
) and acquiring both with the same dose level (11.6 mGy), the pCT appeared to be less noisy than xCT, with an RSP standard deviation of 0.0063. Concerning the RSP accuracy, the measured mean absolute percentage errors were (0.23+-0.09)% in air and (0.21+-0.07)% in water.
. The obtained performances confirm that the INFN pCT system provides a very accurate RSP estimation, appearing to be a feasible clinical tool for verification and correction of xCT calibration in proton treatment planning. |
doi_str_mv | 10.1088/1361-6560/acd6d3 |
format | article |
fullrecord | <record><control><sourceid>proquest_hal_p</sourceid><recordid>TN_cdi_hal_primary_oai_HAL_hal_04255466v1</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2816760503</sourcerecordid><originalsourceid>FETCH-LOGICAL-c445t-4b23aae2b2dcf6e3a11c7efbc3fbc3d9df981dd7fccc2edf4f320ec21507cda73</originalsourceid><addsrcrecordid>eNp9kc1LwzAYh4Mobk7vnqTHCdblo0m74xjODcb0MM8hzQfr6JqadOL8603t3EmEhMCP5_3B-wSAWwQfEcyyESIMxYwyOBJSMUXOQP8UnYM-hATFY0RpD1x5v4UQoQwnl6BHUgwRxeM-eJ1uhBOy0a74Ek1hq8iaqNnoaLGaraLa2SZE03Xkpagq7SJjXSSd9T6Woixyd5r5jJ04BPIaXBhRen1zfAfgbfa0ns7j5cvzYjpZxjJJaBMnOSZCaJxjJQ3TRCAkU21ySdqrxsqMM6RUaqSUWCuTGIKhlhhRmEolUjIA913vRpS8dsVOuAO3ouDzyZK3GUwwpQljHyiww44N-7zvtW_4rvBSl6WotN17jjPEUgYpJAGFHfqzpNPm1I0gb53zVjBvBfPOeRi5O7bv851Wp4FfyQF46IDC1nxr964KYv7rG_6B17ucs0DjcJLwk7xWhnwDq1GYUQ</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2816760503</pqid></control><display><type>article</type><title>Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT</title><source>Institute of Physics</source><creator>Fogazzi, Elena ; Trevisan, Diego ; Farace, Paolo ; Righetto, Roberto ; Rit, Simon ; Scaringella, Monica ; Bruzzi, Mara ; Tommasino, Francesco ; Civinini, Carlo</creator><creatorcontrib>Fogazzi, Elena ; Trevisan, Diego ; Farace, Paolo ; Righetto, Roberto ; Rit, Simon ; Scaringella, Monica ; Bruzzi, Mara ; Tommasino, Francesco ; Civinini, Carlo</creatorcontrib><description>. The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum (NPS) and RSP accuracy has been investigated, as a preliminary step to implement a new cross-calibration method for x-ray CT (xCT).
. The INFN pCT apparatus, made of four planes of silicon micro-strip detectors and a YAG:Ce scintillating calorimeter, reconstructs 3D RSP maps by a filtered-back projection algorithm. The imaging performances (i.e. spatial resolution, NPS and RSP accuracy) of the pCT system were assessed on a custom-made phantom, made of plastic materials with different densities ((0.66, 2.18) g cm
). For comparison, the same phantom was acquired with a clinical xCT system.
. The spatial resolution analysis revealed the nonlinearity of the imaging system, showing different imaging responses in air or water phantom background. Applying the Hann filter in the pCT reconstruction, it was possible to investigate the imaging potential of the system. Matching the spatial resolution value of the xCT (0.54 lp mm
) and acquiring both with the same dose level (11.6 mGy), the pCT appeared to be less noisy than xCT, with an RSP standard deviation of 0.0063. Concerning the RSP accuracy, the measured mean absolute percentage errors were (0.23+-0.09)% in air and (0.21+-0.07)% in water.
. The obtained performances confirm that the INFN pCT system provides a very accurate RSP estimation, appearing to be a feasible clinical tool for verification and correction of xCT calibration in proton treatment planning.</description><identifier>ISSN: 0031-9155</identifier><identifier>EISSN: 1361-6560</identifier><identifier>DOI: 10.1088/1361-6560/acd6d3</identifier><identifier>PMID: 37201529</identifier><identifier>CODEN: PHMBA7</identifier><language>eng</language><publisher>England: IOP Publishing</publisher><subject>Bioengineering ; Calibration ; computed tomography ; image quality ; ion imaging ; Life Sciences ; Medical Physics ; Nuclear medicine ; Phantoms, Imaging ; Physics ; proton therapy ; Protons ; Tomography, X-Ray Computed - methods ; Water ; X-Rays</subject><ispartof>Physics in medicine & biology, 2023-06, Vol.68 (12), p.124001</ispartof><rights>2023 The Author(s). Published on behalf of Institute of Physics and Engineering in Medicine by IOP Publishing Ltd</rights><rights>Creative Commons Attribution license.</rights><rights>Distributed under a Creative Commons Attribution 4.0 International License</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c445t-4b23aae2b2dcf6e3a11c7efbc3fbc3d9df981dd7fccc2edf4f320ec21507cda73</citedby><cites>FETCH-LOGICAL-c445t-4b23aae2b2dcf6e3a11c7efbc3fbc3d9df981dd7fccc2edf4f320ec21507cda73</cites><orcidid>0000-0002-6753-2832 ; 0000-0002-2527-4070 ; 0000-0002-8684-9261 ; 0000-0002-4952-3799 ; 0000-0003-2530-1013</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,778,782,883,27907,27908</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/37201529$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://hal.science/hal-04255466$$DView record in HAL$$Hfree_for_read</backlink></links><search><creatorcontrib>Fogazzi, Elena</creatorcontrib><creatorcontrib>Trevisan, Diego</creatorcontrib><creatorcontrib>Farace, Paolo</creatorcontrib><creatorcontrib>Righetto, Roberto</creatorcontrib><creatorcontrib>Rit, Simon</creatorcontrib><creatorcontrib>Scaringella, Monica</creatorcontrib><creatorcontrib>Bruzzi, Mara</creatorcontrib><creatorcontrib>Tommasino, Francesco</creatorcontrib><creatorcontrib>Civinini, Carlo</creatorcontrib><title>Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT</title><title>Physics in medicine & biology</title><addtitle>PMB</addtitle><addtitle>Phys. Med. Biol</addtitle><description>. The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum (NPS) and RSP accuracy has been investigated, as a preliminary step to implement a new cross-calibration method for x-ray CT (xCT).
. The INFN pCT apparatus, made of four planes of silicon micro-strip detectors and a YAG:Ce scintillating calorimeter, reconstructs 3D RSP maps by a filtered-back projection algorithm. The imaging performances (i.e. spatial resolution, NPS and RSP accuracy) of the pCT system were assessed on a custom-made phantom, made of plastic materials with different densities ((0.66, 2.18) g cm
). For comparison, the same phantom was acquired with a clinical xCT system.
. The spatial resolution analysis revealed the nonlinearity of the imaging system, showing different imaging responses in air or water phantom background. Applying the Hann filter in the pCT reconstruction, it was possible to investigate the imaging potential of the system. Matching the spatial resolution value of the xCT (0.54 lp mm
) and acquiring both with the same dose level (11.6 mGy), the pCT appeared to be less noisy than xCT, with an RSP standard deviation of 0.0063. Concerning the RSP accuracy, the measured mean absolute percentage errors were (0.23+-0.09)% in air and (0.21+-0.07)% in water.
. The obtained performances confirm that the INFN pCT system provides a very accurate RSP estimation, appearing to be a feasible clinical tool for verification and correction of xCT calibration in proton treatment planning.</description><subject>Bioengineering</subject><subject>Calibration</subject><subject>computed tomography</subject><subject>image quality</subject><subject>ion imaging</subject><subject>Life Sciences</subject><subject>Medical Physics</subject><subject>Nuclear medicine</subject><subject>Phantoms, Imaging</subject><subject>Physics</subject><subject>proton therapy</subject><subject>Protons</subject><subject>Tomography, X-Ray Computed - methods</subject><subject>Water</subject><subject>X-Rays</subject><issn>0031-9155</issn><issn>1361-6560</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><recordid>eNp9kc1LwzAYh4Mobk7vnqTHCdblo0m74xjODcb0MM8hzQfr6JqadOL8603t3EmEhMCP5_3B-wSAWwQfEcyyESIMxYwyOBJSMUXOQP8UnYM-hATFY0RpD1x5v4UQoQwnl6BHUgwRxeM-eJ1uhBOy0a74Ek1hq8iaqNnoaLGaraLa2SZE03Xkpagq7SJjXSSd9T6Woixyd5r5jJ04BPIaXBhRen1zfAfgbfa0ns7j5cvzYjpZxjJJaBMnOSZCaJxjJQ3TRCAkU21ySdqrxsqMM6RUaqSUWCuTGIKhlhhRmEolUjIA913vRpS8dsVOuAO3ouDzyZK3GUwwpQljHyiww44N-7zvtW_4rvBSl6WotN17jjPEUgYpJAGFHfqzpNPm1I0gb53zVjBvBfPOeRi5O7bv851Wp4FfyQF46IDC1nxr964KYv7rG_6B17ucs0DjcJLwk7xWhnwDq1GYUQ</recordid><startdate>20230621</startdate><enddate>20230621</enddate><creator>Fogazzi, Elena</creator><creator>Trevisan, Diego</creator><creator>Farace, Paolo</creator><creator>Righetto, Roberto</creator><creator>Rit, Simon</creator><creator>Scaringella, Monica</creator><creator>Bruzzi, Mara</creator><creator>Tommasino, Francesco</creator><creator>Civinini, Carlo</creator><general>IOP Publishing</general><scope>O3W</scope><scope>TSCCA</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>1XC</scope><scope>VOOES</scope><orcidid>https://orcid.org/0000-0002-6753-2832</orcidid><orcidid>https://orcid.org/0000-0002-2527-4070</orcidid><orcidid>https://orcid.org/0000-0002-8684-9261</orcidid><orcidid>https://orcid.org/0000-0002-4952-3799</orcidid><orcidid>https://orcid.org/0000-0003-2530-1013</orcidid></search><sort><creationdate>20230621</creationdate><title>Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT</title><author>Fogazzi, Elena ; Trevisan, Diego ; Farace, Paolo ; Righetto, Roberto ; Rit, Simon ; Scaringella, Monica ; Bruzzi, Mara ; Tommasino, Francesco ; Civinini, Carlo</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c445t-4b23aae2b2dcf6e3a11c7efbc3fbc3d9df981dd7fccc2edf4f320ec21507cda73</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Bioengineering</topic><topic>Calibration</topic><topic>computed tomography</topic><topic>image quality</topic><topic>ion imaging</topic><topic>Life Sciences</topic><topic>Medical Physics</topic><topic>Nuclear medicine</topic><topic>Phantoms, Imaging</topic><topic>Physics</topic><topic>proton therapy</topic><topic>Protons</topic><topic>Tomography, X-Ray Computed - methods</topic><topic>Water</topic><topic>X-Rays</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fogazzi, Elena</creatorcontrib><creatorcontrib>Trevisan, Diego</creatorcontrib><creatorcontrib>Farace, Paolo</creatorcontrib><creatorcontrib>Righetto, Roberto</creatorcontrib><creatorcontrib>Rit, Simon</creatorcontrib><creatorcontrib>Scaringella, Monica</creatorcontrib><creatorcontrib>Bruzzi, Mara</creatorcontrib><creatorcontrib>Tommasino, Francesco</creatorcontrib><creatorcontrib>Civinini, Carlo</creatorcontrib><collection>Open Access: IOP Publishing Free Content</collection><collection>IOPscience (Open Access)</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Hyper Article en Ligne (HAL)</collection><collection>Hyper Article en Ligne (HAL) (Open Access)</collection><jtitle>Physics in medicine & biology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fogazzi, Elena</au><au>Trevisan, Diego</au><au>Farace, Paolo</au><au>Righetto, Roberto</au><au>Rit, Simon</au><au>Scaringella, Monica</au><au>Bruzzi, Mara</au><au>Tommasino, Francesco</au><au>Civinini, Carlo</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT</atitle><jtitle>Physics in medicine & biology</jtitle><stitle>PMB</stitle><addtitle>Phys. Med. Biol</addtitle><date>2023-06-21</date><risdate>2023</risdate><volume>68</volume><issue>12</issue><spage>124001</spage><pages>124001-</pages><issn>0031-9155</issn><eissn>1361-6560</eissn><coden>PHMBA7</coden><abstract>. The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum (NPS) and RSP accuracy has been investigated, as a preliminary step to implement a new cross-calibration method for x-ray CT (xCT).
. The INFN pCT apparatus, made of four planes of silicon micro-strip detectors and a YAG:Ce scintillating calorimeter, reconstructs 3D RSP maps by a filtered-back projection algorithm. The imaging performances (i.e. spatial resolution, NPS and RSP accuracy) of the pCT system were assessed on a custom-made phantom, made of plastic materials with different densities ((0.66, 2.18) g cm
). For comparison, the same phantom was acquired with a clinical xCT system.
. The spatial resolution analysis revealed the nonlinearity of the imaging system, showing different imaging responses in air or water phantom background. Applying the Hann filter in the pCT reconstruction, it was possible to investigate the imaging potential of the system. Matching the spatial resolution value of the xCT (0.54 lp mm
) and acquiring both with the same dose level (11.6 mGy), the pCT appeared to be less noisy than xCT, with an RSP standard deviation of 0.0063. Concerning the RSP accuracy, the measured mean absolute percentage errors were (0.23+-0.09)% in air and (0.21+-0.07)% in water.
. The obtained performances confirm that the INFN pCT system provides a very accurate RSP estimation, appearing to be a feasible clinical tool for verification and correction of xCT calibration in proton treatment planning.</abstract><cop>England</cop><pub>IOP Publishing</pub><pmid>37201529</pmid><doi>10.1088/1361-6560/acd6d3</doi><tpages>13</tpages><orcidid>https://orcid.org/0000-0002-6753-2832</orcidid><orcidid>https://orcid.org/0000-0002-2527-4070</orcidid><orcidid>https://orcid.org/0000-0002-8684-9261</orcidid><orcidid>https://orcid.org/0000-0002-4952-3799</orcidid><orcidid>https://orcid.org/0000-0003-2530-1013</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0031-9155 |
ispartof | Physics in medicine & biology, 2023-06, Vol.68 (12), p.124001 |
issn | 0031-9155 1361-6560 |
language | eng |
recordid | cdi_hal_primary_oai_HAL_hal_04255466v1 |
source | Institute of Physics |
subjects | Bioengineering Calibration computed tomography image quality ion imaging Life Sciences Medical Physics Nuclear medicine Phantoms, Imaging Physics proton therapy Protons Tomography, X-Ray Computed - methods Water X-Rays |
title | Characterization of the INFN proton CT scanner for cross-calibration of x-ray CT |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-16T14%3A37%3A09IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_hal_p&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Characterization%20of%20the%20INFN%20proton%20CT%20scanner%20for%20cross-calibration%20of%20x-ray%20CT&rft.jtitle=Physics%20in%20medicine%20&%20biology&rft.au=Fogazzi,%20Elena&rft.date=2023-06-21&rft.volume=68&rft.issue=12&rft.spage=124001&rft.pages=124001-&rft.issn=0031-9155&rft.eissn=1361-6560&rft.coden=PHMBA7&rft_id=info:doi/10.1088/1361-6560/acd6d3&rft_dat=%3Cproquest_hal_p%3E2816760503%3C/proquest_hal_p%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c445t-4b23aae2b2dcf6e3a11c7efbc3fbc3d9df981dd7fccc2edf4f320ec21507cda73%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2816760503&rft_id=info:pmid/37201529&rfr_iscdi=true |