Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer
Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer wi...
Saved in:
Published in: | Physica medica 2024-01, Vol.117, p.103203-103203, Article 103203 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | |
---|---|
cites | cdi_FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3 |
container_end_page | 103203 |
container_issue | |
container_start_page | 103203 |
container_title | Physica medica |
container_volume | 117 |
creator | Heikkilä, Annele Boman, Eeva Rossi, Maija Vanhanen, Antti Mankinen, Mikko Postema, Michiel Koivumäki, Tuomas |
description | Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
•Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors. |
doi_str_mv | 10.1016/j.ejmp.2023.103203 |
format | article |
fullrecord | <record><control><sourceid>proquest_hal_p</sourceid><recordid>TN_cdi_hal_primary_oai_HAL_hal_04366710v1</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S1120179723012322</els_id><sourcerecordid>2910195954</sourcerecordid><originalsourceid>FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3</originalsourceid><addsrcrecordid>eNp9kc9q3DAQh01padK0L9BD0bE9eKOx_E_QS0japrDQSwK9iVlpRLTYlivJC3mEvnW08TbHgkDD8M0PZr6i-Ah8Axzay_2G9uO8qXglckNUXLwqzqGr6hIk_H6da6h4CZ3szop3Me55ZqqmeVuciR46qECeF39vfHQjpeA0I2tJJ-YtCz5hcn7CgUVKy8woBB8icxM7-GE58aM3y4CJDMOgWXqggPMjw8kw62gwpZvK54KlQJhGmp6zB7KpjM7ksV3ux8Q0TprC--KNxSHSh9N_Udx__3Z3fVtuf_34eX21LbXom5RX24laY2-Q6rbVRlogvWtBNLIGK3XbN5IbAdBmviONtkcpasyvrZG0uCi-rLkPOKg5uBHDo_Lo1O3VVh17vBZt2wE_QGY_r-wc_J-FYlKji5qGASfyS1SVzCJkI5s6o9WK6uBjDGRfsoGroy61V0dd6qhLrbry0KdT_rIbybyM_POTga8rQPkiB0dBRe0on8u4kF0p493_8p8AaeuoIA</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2910195954</pqid></control><display><type>article</type><title>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</title><source>ScienceDirect Freedom Collection</source><creator>Heikkilä, Annele ; Boman, Eeva ; Rossi, Maija ; Vanhanen, Antti ; Mankinen, Mikko ; Postema, Michiel ; Koivumäki, Tuomas</creator><creatorcontrib>Heikkilä, Annele ; Boman, Eeva ; Rossi, Maija ; Vanhanen, Antti ; Mankinen, Mikko ; Postema, Michiel ; Koivumäki, Tuomas</creatorcontrib><description>Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
•Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors.</description><identifier>ISSN: 1120-1797</identifier><identifier>EISSN: 1724-191X</identifier><identifier>DOI: 10.1016/j.ejmp.2023.103203</identifier><identifier>PMID: 38171219</identifier><language>eng</language><publisher>Italy: Elsevier Ltd</publisher><subject>Breast radiotherapy ; DICOM image rotation ; Interfraction positioning errors ; Medical Physics ; Physics ; Robustness of VMAT</subject><ispartof>Physica medica, 2024-01, Vol.117, p.103203-103203, Article 103203</ispartof><rights>2023 Associazione Italiana di Fisica Medica e Sanitaria</rights><rights>Copyright © 2023 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.</rights><rights>Attribution</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3</cites><orcidid>0000-0002-7920-2531 ; 0000-0002-2266-1438 ; 0000-0001-5887-1739 ; 0000-0001-5639-8765</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,777,781,882,27905,27906</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/38171219$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://hal.science/hal-04366710$$DView record in HAL$$Hfree_for_read</backlink></links><search><creatorcontrib>Heikkilä, Annele</creatorcontrib><creatorcontrib>Boman, Eeva</creatorcontrib><creatorcontrib>Rossi, Maija</creatorcontrib><creatorcontrib>Vanhanen, Antti</creatorcontrib><creatorcontrib>Mankinen, Mikko</creatorcontrib><creatorcontrib>Postema, Michiel</creatorcontrib><creatorcontrib>Koivumäki, Tuomas</creatorcontrib><title>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</title><title>Physica medica</title><addtitle>Phys Med</addtitle><description>Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
•Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors.</description><subject>Breast radiotherapy</subject><subject>DICOM image rotation</subject><subject>Interfraction positioning errors</subject><subject>Medical Physics</subject><subject>Physics</subject><subject>Robustness of VMAT</subject><issn>1120-1797</issn><issn>1724-191X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2024</creationdate><recordtype>article</recordtype><recordid>eNp9kc9q3DAQh01padK0L9BD0bE9eKOx_E_QS0japrDQSwK9iVlpRLTYlivJC3mEvnW08TbHgkDD8M0PZr6i-Ah8Axzay_2G9uO8qXglckNUXLwqzqGr6hIk_H6da6h4CZ3szop3Me55ZqqmeVuciR46qECeF39vfHQjpeA0I2tJJ-YtCz5hcn7CgUVKy8woBB8icxM7-GE58aM3y4CJDMOgWXqggPMjw8kw62gwpZvK54KlQJhGmp6zB7KpjM7ksV3ux8Q0TprC--KNxSHSh9N_Udx__3Z3fVtuf_34eX21LbXom5RX24laY2-Q6rbVRlogvWtBNLIGK3XbN5IbAdBmviONtkcpasyvrZG0uCi-rLkPOKg5uBHDo_Lo1O3VVh17vBZt2wE_QGY_r-wc_J-FYlKji5qGASfyS1SVzCJkI5s6o9WK6uBjDGRfsoGroy61V0dd6qhLrbry0KdT_rIbybyM_POTga8rQPkiB0dBRe0on8u4kF0p493_8p8AaeuoIA</recordid><startdate>20240101</startdate><enddate>20240101</enddate><creator>Heikkilä, Annele</creator><creator>Boman, Eeva</creator><creator>Rossi, Maija</creator><creator>Vanhanen, Antti</creator><creator>Mankinen, Mikko</creator><creator>Postema, Michiel</creator><creator>Koivumäki, Tuomas</creator><general>Elsevier Ltd</general><general>Elsevier</general><scope>6I.</scope><scope>AAFTH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>1XC</scope><scope>VOOES</scope><orcidid>https://orcid.org/0000-0002-7920-2531</orcidid><orcidid>https://orcid.org/0000-0002-2266-1438</orcidid><orcidid>https://orcid.org/0000-0001-5887-1739</orcidid><orcidid>https://orcid.org/0000-0001-5639-8765</orcidid></search><sort><creationdate>20240101</creationdate><title>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</title><author>Heikkilä, Annele ; Boman, Eeva ; Rossi, Maija ; Vanhanen, Antti ; Mankinen, Mikko ; Postema, Michiel ; Koivumäki, Tuomas</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2024</creationdate><topic>Breast radiotherapy</topic><topic>DICOM image rotation</topic><topic>Interfraction positioning errors</topic><topic>Medical Physics</topic><topic>Physics</topic><topic>Robustness of VMAT</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Heikkilä, Annele</creatorcontrib><creatorcontrib>Boman, Eeva</creatorcontrib><creatorcontrib>Rossi, Maija</creatorcontrib><creatorcontrib>Vanhanen, Antti</creatorcontrib><creatorcontrib>Mankinen, Mikko</creatorcontrib><creatorcontrib>Postema, Michiel</creatorcontrib><creatorcontrib>Koivumäki, Tuomas</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Hyper Article en Ligne (HAL)</collection><collection>Hyper Article en Ligne (HAL) (Open Access)</collection><jtitle>Physica medica</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Heikkilä, Annele</au><au>Boman, Eeva</au><au>Rossi, Maija</au><au>Vanhanen, Antti</au><au>Mankinen, Mikko</au><au>Postema, Michiel</au><au>Koivumäki, Tuomas</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</atitle><jtitle>Physica medica</jtitle><addtitle>Phys Med</addtitle><date>2024-01-01</date><risdate>2024</risdate><volume>117</volume><spage>103203</spage><epage>103203</epage><pages>103203-103203</pages><artnum>103203</artnum><issn>1120-1797</issn><eissn>1724-191X</eissn><abstract>Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
•Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors.</abstract><cop>Italy</cop><pub>Elsevier Ltd</pub><pmid>38171219</pmid><doi>10.1016/j.ejmp.2023.103203</doi><tpages>1</tpages><orcidid>https://orcid.org/0000-0002-7920-2531</orcidid><orcidid>https://orcid.org/0000-0002-2266-1438</orcidid><orcidid>https://orcid.org/0000-0001-5887-1739</orcidid><orcidid>https://orcid.org/0000-0001-5639-8765</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1120-1797 |
ispartof | Physica medica, 2024-01, Vol.117, p.103203-103203, Article 103203 |
issn | 1120-1797 1724-191X |
language | eng |
recordid | cdi_hal_primary_oai_HAL_hal_04366710v1 |
source | ScienceDirect Freedom Collection |
subjects | Breast radiotherapy DICOM image rotation Interfraction positioning errors Medical Physics Physics Robustness of VMAT |
title | Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-19T23%3A41%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_hal_p&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Dosimetric%20effect%20of%20rotational%20setup%20errors%20in%20volumetric%20modulated%20arc%20therapy%20and%20field-in-field%20treatment%20of%20left-sided%20breast%20cancer&rft.jtitle=Physica%20medica&rft.au=Heikkil%C3%A4,%20Annele&rft.date=2024-01-01&rft.volume=117&rft.spage=103203&rft.epage=103203&rft.pages=103203-103203&rft.artnum=103203&rft.issn=1120-1797&rft.eissn=1724-191X&rft_id=info:doi/10.1016/j.ejmp.2023.103203&rft_dat=%3Cproquest_hal_p%3E2910195954%3C/proquest_hal_p%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2910195954&rft_id=info:pmid/38171219&rfr_iscdi=true |