Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer

Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer wi...

Full description

Saved in:
Bibliographic Details
Published in:Physica medica 2024-01, Vol.117, p.103203-103203, Article 103203
Main Authors: Heikkilä, Annele, Boman, Eeva, Rossi, Maija, Vanhanen, Antti, Mankinen, Mikko, Postema, Michiel, Koivumäki, Tuomas
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites cdi_FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3
container_end_page 103203
container_issue
container_start_page 103203
container_title Physica medica
container_volume 117
creator Heikkilä, Annele
Boman, Eeva
Rossi, Maija
Vanhanen, Antti
Mankinen, Mikko
Postema, Michiel
Koivumäki, Tuomas
description Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors. •Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors.
doi_str_mv 10.1016/j.ejmp.2023.103203
format article
fullrecord <record><control><sourceid>proquest_hal_p</sourceid><recordid>TN_cdi_hal_primary_oai_HAL_hal_04366710v1</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S1120179723012322</els_id><sourcerecordid>2910195954</sourcerecordid><originalsourceid>FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3</originalsourceid><addsrcrecordid>eNp9kc9q3DAQh01padK0L9BD0bE9eKOx_E_QS0japrDQSwK9iVlpRLTYlivJC3mEvnW08TbHgkDD8M0PZr6i-Ah8Axzay_2G9uO8qXglckNUXLwqzqGr6hIk_H6da6h4CZ3szop3Me55ZqqmeVuciR46qECeF39vfHQjpeA0I2tJJ-YtCz5hcn7CgUVKy8woBB8icxM7-GE58aM3y4CJDMOgWXqggPMjw8kw62gwpZvK54KlQJhGmp6zB7KpjM7ksV3ux8Q0TprC--KNxSHSh9N_Udx__3Z3fVtuf_34eX21LbXom5RX24laY2-Q6rbVRlogvWtBNLIGK3XbN5IbAdBmviONtkcpasyvrZG0uCi-rLkPOKg5uBHDo_Lo1O3VVh17vBZt2wE_QGY_r-wc_J-FYlKji5qGASfyS1SVzCJkI5s6o9WK6uBjDGRfsoGroy61V0dd6qhLrbry0KdT_rIbybyM_POTga8rQPkiB0dBRe0on8u4kF0p493_8p8AaeuoIA</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2910195954</pqid></control><display><type>article</type><title>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</title><source>ScienceDirect Freedom Collection</source><creator>Heikkilä, Annele ; Boman, Eeva ; Rossi, Maija ; Vanhanen, Antti ; Mankinen, Mikko ; Postema, Michiel ; Koivumäki, Tuomas</creator><creatorcontrib>Heikkilä, Annele ; Boman, Eeva ; Rossi, Maija ; Vanhanen, Antti ; Mankinen, Mikko ; Postema, Michiel ; Koivumäki, Tuomas</creatorcontrib><description>Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors. •Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors.</description><identifier>ISSN: 1120-1797</identifier><identifier>EISSN: 1724-191X</identifier><identifier>DOI: 10.1016/j.ejmp.2023.103203</identifier><identifier>PMID: 38171219</identifier><language>eng</language><publisher>Italy: Elsevier Ltd</publisher><subject>Breast radiotherapy ; DICOM image rotation ; Interfraction positioning errors ; Medical Physics ; Physics ; Robustness of VMAT</subject><ispartof>Physica medica, 2024-01, Vol.117, p.103203-103203, Article 103203</ispartof><rights>2023 Associazione Italiana di Fisica Medica e Sanitaria</rights><rights>Copyright © 2023 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.</rights><rights>Attribution</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3</cites><orcidid>0000-0002-7920-2531 ; 0000-0002-2266-1438 ; 0000-0001-5887-1739 ; 0000-0001-5639-8765</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,777,781,882,27905,27906</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/38171219$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://hal.science/hal-04366710$$DView record in HAL$$Hfree_for_read</backlink></links><search><creatorcontrib>Heikkilä, Annele</creatorcontrib><creatorcontrib>Boman, Eeva</creatorcontrib><creatorcontrib>Rossi, Maija</creatorcontrib><creatorcontrib>Vanhanen, Antti</creatorcontrib><creatorcontrib>Mankinen, Mikko</creatorcontrib><creatorcontrib>Postema, Michiel</creatorcontrib><creatorcontrib>Koivumäki, Tuomas</creatorcontrib><title>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</title><title>Physica medica</title><addtitle>Phys Med</addtitle><description>Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors. •Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors.</description><subject>Breast radiotherapy</subject><subject>DICOM image rotation</subject><subject>Interfraction positioning errors</subject><subject>Medical Physics</subject><subject>Physics</subject><subject>Robustness of VMAT</subject><issn>1120-1797</issn><issn>1724-191X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2024</creationdate><recordtype>article</recordtype><recordid>eNp9kc9q3DAQh01padK0L9BD0bE9eKOx_E_QS0japrDQSwK9iVlpRLTYlivJC3mEvnW08TbHgkDD8M0PZr6i-Ah8Axzay_2G9uO8qXglckNUXLwqzqGr6hIk_H6da6h4CZ3szop3Me55ZqqmeVuciR46qECeF39vfHQjpeA0I2tJJ-YtCz5hcn7CgUVKy8woBB8icxM7-GE58aM3y4CJDMOgWXqggPMjw8kw62gwpZvK54KlQJhGmp6zB7KpjM7ksV3ux8Q0TprC--KNxSHSh9N_Udx__3Z3fVtuf_34eX21LbXom5RX24laY2-Q6rbVRlogvWtBNLIGK3XbN5IbAdBmviONtkcpasyvrZG0uCi-rLkPOKg5uBHDo_Lo1O3VVh17vBZt2wE_QGY_r-wc_J-FYlKji5qGASfyS1SVzCJkI5s6o9WK6uBjDGRfsoGroy61V0dd6qhLrbry0KdT_rIbybyM_POTga8rQPkiB0dBRe0on8u4kF0p493_8p8AaeuoIA</recordid><startdate>20240101</startdate><enddate>20240101</enddate><creator>Heikkilä, Annele</creator><creator>Boman, Eeva</creator><creator>Rossi, Maija</creator><creator>Vanhanen, Antti</creator><creator>Mankinen, Mikko</creator><creator>Postema, Michiel</creator><creator>Koivumäki, Tuomas</creator><general>Elsevier Ltd</general><general>Elsevier</general><scope>6I.</scope><scope>AAFTH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>1XC</scope><scope>VOOES</scope><orcidid>https://orcid.org/0000-0002-7920-2531</orcidid><orcidid>https://orcid.org/0000-0002-2266-1438</orcidid><orcidid>https://orcid.org/0000-0001-5887-1739</orcidid><orcidid>https://orcid.org/0000-0001-5639-8765</orcidid></search><sort><creationdate>20240101</creationdate><title>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</title><author>Heikkilä, Annele ; Boman, Eeva ; Rossi, Maija ; Vanhanen, Antti ; Mankinen, Mikko ; Postema, Michiel ; Koivumäki, Tuomas</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2024</creationdate><topic>Breast radiotherapy</topic><topic>DICOM image rotation</topic><topic>Interfraction positioning errors</topic><topic>Medical Physics</topic><topic>Physics</topic><topic>Robustness of VMAT</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Heikkilä, Annele</creatorcontrib><creatorcontrib>Boman, Eeva</creatorcontrib><creatorcontrib>Rossi, Maija</creatorcontrib><creatorcontrib>Vanhanen, Antti</creatorcontrib><creatorcontrib>Mankinen, Mikko</creatorcontrib><creatorcontrib>Postema, Michiel</creatorcontrib><creatorcontrib>Koivumäki, Tuomas</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Hyper Article en Ligne (HAL)</collection><collection>Hyper Article en Ligne (HAL) (Open Access)</collection><jtitle>Physica medica</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Heikkilä, Annele</au><au>Boman, Eeva</au><au>Rossi, Maija</au><au>Vanhanen, Antti</au><au>Mankinen, Mikko</au><au>Postema, Michiel</au><au>Koivumäki, Tuomas</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer</atitle><jtitle>Physica medica</jtitle><addtitle>Phys Med</addtitle><date>2024-01-01</date><risdate>2024</risdate><volume>117</volume><spage>103203</spage><epage>103203</epage><pages>103203-103203</pages><artnum>103203</artnum><issn>1120-1797</issn><eissn>1724-191X</eissn><abstract>Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors. •Rotational setup errors were simulated by rotating CT images using a Matlab script.•Clinical target volume dose distribution was robust towards setup errors up to 3°.•Heart and LAD doses were more robust in VMAT plans compared to field-in-field plans.•Intervention tolerance of 1° is recommended for rotational setup errors.</abstract><cop>Italy</cop><pub>Elsevier Ltd</pub><pmid>38171219</pmid><doi>10.1016/j.ejmp.2023.103203</doi><tpages>1</tpages><orcidid>https://orcid.org/0000-0002-7920-2531</orcidid><orcidid>https://orcid.org/0000-0002-2266-1438</orcidid><orcidid>https://orcid.org/0000-0001-5887-1739</orcidid><orcidid>https://orcid.org/0000-0001-5639-8765</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1120-1797
ispartof Physica medica, 2024-01, Vol.117, p.103203-103203, Article 103203
issn 1120-1797
1724-191X
language eng
recordid cdi_hal_primary_oai_HAL_hal_04366710v1
source ScienceDirect Freedom Collection
subjects Breast radiotherapy
DICOM image rotation
Interfraction positioning errors
Medical Physics
Physics
Robustness of VMAT
title Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-19T23%3A41%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_hal_p&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Dosimetric%20effect%20of%20rotational%20setup%20errors%20in%20volumetric%20modulated%20arc%20therapy%20and%20field-in-field%20treatment%20of%20left-sided%20breast%20cancer&rft.jtitle=Physica%20medica&rft.au=Heikkil%C3%A4,%20Annele&rft.date=2024-01-01&rft.volume=117&rft.spage=103203&rft.epage=103203&rft.pages=103203-103203&rft.artnum=103203&rft.issn=1120-1797&rft.eissn=1724-191X&rft_id=info:doi/10.1016/j.ejmp.2023.103203&rft_dat=%3Cproquest_hal_p%3E2910195954%3C/proquest_hal_p%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c385t-19b34ca8dae466cd9f1ecb6135941f9c68590d3116c387ecaf8a934a34a64aec3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2910195954&rft_id=info:pmid/38171219&rfr_iscdi=true