Multimodality imaging guidance for percutaneous paravalvular leak closure: Insights from the multi-centre FFPP register
Percutaneous paravalvular leak (PVL) closure has emerged as a palliative alternative to surgical management in selected high-risk patients. Percutaneous procedures are challenging, especially for mitral PVL. Accurate imaging of the morphologies of the defects is mandatory, together with precise guid...
Saved in:
Published in: | Archives of cardiovascular diseases 2018-06, Vol.111 (6-7), p.421-431 |
---|---|
Main Authors: | , , , , , , , , , , , , , , , , , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-c434t-83226436ccd4c44f696cfe5862d93b5a4279e9b41a204ac89f4bbe484274f0623 |
---|---|
cites | cdi_FETCH-LOGICAL-c434t-83226436ccd4c44f696cfe5862d93b5a4279e9b41a204ac89f4bbe484274f0623 |
container_end_page | 431 |
container_issue | 6-7 |
container_start_page | 421 |
container_title | Archives of cardiovascular diseases |
container_volume | 111 |
creator | Hascoet, Sébastien Smolka, Grzegorz Bagate, Francois Guihaire, Julien Potier, Agathe Hadeed, Khaled Lavie-Badie, Yoan Bouvaist, Hélène Dauphin, Claire Bauer, Fabrice Nejjari, Mohammed Pillière, Remy Brochet, Eric Mangin, Lionel Bonnet, Guillaume Ciobotaru, Vlad Leurent, Guillaume Hammoudi, Najib Aminian, Adel Karsenty, Clement Spaulding, Christian Armero, Sebastien Collet, Frederic Champagnac, Didier Ternacle, Julien Kloeckner, Martin Gerardin, Benoit Isorni, Marc-Antoine |
description | Percutaneous paravalvular leak (PVL) closure has emerged as a palliative alternative to surgical management in selected high-risk patients. Percutaneous procedures are challenging, especially for mitral PVL. Accurate imaging of the morphologies of the defects is mandatory, together with precise guidance in the catheterization laboratory to enhance success rates.
To describe imaging modalities used in clinical practice to guide percutaneous PVL closure and assess the potential of new imaging tools.
Data from the ‘Fermeture de Fuite paraprothétique’ (FFPP) register were used. The FFPP register is an international multi-institutional collaborative register started in 2017 with a retrospective and a prospective part. A descriptive analysis of multimodality imaging used to guide PVL closure in clinical practice was performed.
Data from 173 procedures performed in 19 centres from three countries (France, Belgium and Poland) were collected, which included eight cases of PVL following transcatheter valve replacement. Transoesophageal echocardiography was used in 167 cases (96.5%) and 3D echocardiography in 87.4% of cases. In one case, 3D-echocardiography was fused with fluoroscopy images in real time using echonavigator software. Details about multimodality imaging were available from a sample of 31 patients. Cardiac computed tomography (CT) was performed before 10 of the procedures. In one case, fusion between preprocedural cardiac CT angiography data and fluoroscopy data was used. In two cases, a 3D model of the valve with PVL was printed.
Echocardiography, particularly the 3D mode, is the cornerstone of PVL imaging. Other imaging modalities, such as cardiac CT and cardiac magnetic resonance imaging, may be of complementary interest. New techniques such as imaging fusion and printing may further facilitate the percutaneous approach of PVLs.
La fermeture percutanée des fuites valvulaires paraprothétiques (FPP) est devenue une alternative à la chirurgie chez des patients sélectionnés à haut-risque. Les procédures sont complexes en particulier pour les FPP mitrales. Une imagerie de précision est nécessaire pour décrire la morphologie des défects, guider la procédure percutanée et augmenter le taux de succès.
L’objectif de cette étude est de décrire les différents examens d’imagerie utilisés en pratique courante et leur intérêt respectifs ainsi que l’éventuel apport des nouvelles techniques.
Les données du registre « fermeture de fuite paraprothétique » (FFPP) ont é |
doi_str_mv | 10.1016/j.acvd.2018.05.001 |
format | article |
fullrecord | <record><control><sourceid>proquest_hal_p</sourceid><recordid>TN_cdi_hal_primary_oai_HAL_hal_04395214v1</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S1875213618300937</els_id><sourcerecordid>2059043677</sourcerecordid><originalsourceid>FETCH-LOGICAL-c434t-83226436ccd4c44f696cfe5862d93b5a4279e9b41a204ac89f4bbe484274f0623</originalsourceid><addsrcrecordid>eNp9kcFu1DAURSMEoqXwAyyQl7BIsB3HiRGbqmLaSoPoAtaW47xkPDjxYDtB_XscpZ0lK1tP513pvpNl7wkuCCb887FQeukKiklT4KrAmLzILklTVzkltHl5_pf8InsTwhFjTuuav84uqBBlTXF9mf39PttoRtcpa-IjMqMazDSgYTadmjSg3nl0Aq_nqCZwc0An5dWi7DJb5ZEF9Rtp68Ls4Qu6n4IZDjGg3rsRxQOgcQ3PNUzRA9rtHh6Qh8GECP5t9qpXNsC7p_cq-7X79vPmLt__uL2_ud7nmpUs5k1JKWcl17pjmrGeC657qBpOO1G2lWK0FiBaRhTFTOlG9KxtgTVpzvpUt7zKPm25B2Xlyad-_lE6ZeTd9V6uM8xKUVHCFpLYjxt78u7PDCHK0QQN1m7VJcWVSDiv64TSDdXeheChP2cTLFc58ihXOXKVI3Elk5y09OEpf25H6M4rzzYS8HUDIF1kMeBl0AaShs540FF2zvwv_x9xaaD7</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2059043677</pqid></control><display><type>article</type><title>Multimodality imaging guidance for percutaneous paravalvular leak closure: Insights from the multi-centre FFPP register</title><source>ScienceDirect Freedom Collection</source><creator>Hascoet, Sébastien ; Smolka, Grzegorz ; Bagate, Francois ; Guihaire, Julien ; Potier, Agathe ; Hadeed, Khaled ; Lavie-Badie, Yoan ; Bouvaist, Hélène ; Dauphin, Claire ; Bauer, Fabrice ; Nejjari, Mohammed ; Pillière, Remy ; Brochet, Eric ; Mangin, Lionel ; Bonnet, Guillaume ; Ciobotaru, Vlad ; Leurent, Guillaume ; Hammoudi, Najib ; Aminian, Adel ; Karsenty, Clement ; Spaulding, Christian ; Armero, Sebastien ; Collet, Frederic ; Champagnac, Didier ; Ternacle, Julien ; Kloeckner, Martin ; Gerardin, Benoit ; Isorni, Marc-Antoine</creator><creatorcontrib>Hascoet, Sébastien ; Smolka, Grzegorz ; Bagate, Francois ; Guihaire, Julien ; Potier, Agathe ; Hadeed, Khaled ; Lavie-Badie, Yoan ; Bouvaist, Hélène ; Dauphin, Claire ; Bauer, Fabrice ; Nejjari, Mohammed ; Pillière, Remy ; Brochet, Eric ; Mangin, Lionel ; Bonnet, Guillaume ; Ciobotaru, Vlad ; Leurent, Guillaume ; Hammoudi, Najib ; Aminian, Adel ; Karsenty, Clement ; Spaulding, Christian ; Armero, Sebastien ; Collet, Frederic ; Champagnac, Didier ; Ternacle, Julien ; Kloeckner, Martin ; Gerardin, Benoit ; Isorni, Marc-Antoine</creatorcontrib><description>Percutaneous paravalvular leak (PVL) closure has emerged as a palliative alternative to surgical management in selected high-risk patients. Percutaneous procedures are challenging, especially for mitral PVL. Accurate imaging of the morphologies of the defects is mandatory, together with precise guidance in the catheterization laboratory to enhance success rates.
To describe imaging modalities used in clinical practice to guide percutaneous PVL closure and assess the potential of new imaging tools.
Data from the ‘Fermeture de Fuite paraprothétique’ (FFPP) register were used. The FFPP register is an international multi-institutional collaborative register started in 2017 with a retrospective and a prospective part. A descriptive analysis of multimodality imaging used to guide PVL closure in clinical practice was performed.
Data from 173 procedures performed in 19 centres from three countries (France, Belgium and Poland) were collected, which included eight cases of PVL following transcatheter valve replacement. Transoesophageal echocardiography was used in 167 cases (96.5%) and 3D echocardiography in 87.4% of cases. In one case, 3D-echocardiography was fused with fluoroscopy images in real time using echonavigator software. Details about multimodality imaging were available from a sample of 31 patients. Cardiac computed tomography (CT) was performed before 10 of the procedures. In one case, fusion between preprocedural cardiac CT angiography data and fluoroscopy data was used. In two cases, a 3D model of the valve with PVL was printed.
Echocardiography, particularly the 3D mode, is the cornerstone of PVL imaging. Other imaging modalities, such as cardiac CT and cardiac magnetic resonance imaging, may be of complementary interest. New techniques such as imaging fusion and printing may further facilitate the percutaneous approach of PVLs.
La fermeture percutanée des fuites valvulaires paraprothétiques (FPP) est devenue une alternative à la chirurgie chez des patients sélectionnés à haut-risque. Les procédures sont complexes en particulier pour les FPP mitrales. Une imagerie de précision est nécessaire pour décrire la morphologie des défects, guider la procédure percutanée et augmenter le taux de succès.
L’objectif de cette étude est de décrire les différents examens d’imagerie utilisés en pratique courante et leur intérêt respectifs ainsi que l’éventuel apport des nouvelles techniques.
Les données du registre « fermeture de fuite paraprothétique » (FFPP) ont été utilisées. Ce registre multicentrique international débuté en 2017 comporte un bras rétrospectif et un bras prospectif. Les examens d’imagerie utilisés pour la fermeture percutanée des FPP ont été décrits.
L’étude a inclus 173 procédures réalisées dans 19 centres de trois pays, incluant huit procédures sur des FPP sur valves implantées par voie percutanée. L’échographie transoesophagienne a été utilisée dans 167 cas (96,5 %) dont 87,4 % avec le mode 3D. Dans un cas, l’échocardiographie 3D a été fusionnée avec la fluoroscopie en temps réel. Au sein d’un échantillon de 31 patients dont les données des différents examens d’imagerie étaient disponibles, 10 scanners cardiaques ont été réalisés. Dans un cas, ce scanner a été fusionné avec la fluoroscopie durant la procédure. Deux modèles 3D de valves avec FPP ont été imprimés.
L’échocardiographie et en particulier le mode 3D est la pierre angulaire de l’imagerie des FPP. Les autres techniques d’imagerie incluant le scanner et l’IRM sont complémentaires. Les nouvelles techniques incluant la fusion et l’impression 3D contribueront possiblement à mieux appréhender les FPP et leur traitement percutané.</description><identifier>ISSN: 1875-2136</identifier><identifier>EISSN: 1875-2128</identifier><identifier>DOI: 10.1016/j.acvd.2018.05.001</identifier><identifier>PMID: 29937207</identifier><language>eng</language><publisher>Netherlands: Elsevier Masson SAS</publisher><subject>3D echocardiography ; 4D-flow MRI ; Cardiac valve ; Echocardiographie 3D ; Fuite paraprothétique ; IRM 4D flow ; Life Sciences ; Paravalvular leak ; Valve cardiaque</subject><ispartof>Archives of cardiovascular diseases, 2018-06, Vol.111 (6-7), p.421-431</ispartof><rights>2018 Elsevier Masson SAS</rights><rights>Copyright © 2018 Elsevier Masson SAS. All rights reserved.</rights><rights>Distributed under a Creative Commons Attribution 4.0 International License</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c434t-83226436ccd4c44f696cfe5862d93b5a4279e9b41a204ac89f4bbe484274f0623</citedby><cites>FETCH-LOGICAL-c434t-83226436ccd4c44f696cfe5862d93b5a4279e9b41a204ac89f4bbe484274f0623</cites><orcidid>0000-0001-6506-1224</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29937207$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://hal.u-pec.fr/hal-04395214$$DView record in HAL$$Hfree_for_read</backlink></links><search><creatorcontrib>Hascoet, Sébastien</creatorcontrib><creatorcontrib>Smolka, Grzegorz</creatorcontrib><creatorcontrib>Bagate, Francois</creatorcontrib><creatorcontrib>Guihaire, Julien</creatorcontrib><creatorcontrib>Potier, Agathe</creatorcontrib><creatorcontrib>Hadeed, Khaled</creatorcontrib><creatorcontrib>Lavie-Badie, Yoan</creatorcontrib><creatorcontrib>Bouvaist, Hélène</creatorcontrib><creatorcontrib>Dauphin, Claire</creatorcontrib><creatorcontrib>Bauer, Fabrice</creatorcontrib><creatorcontrib>Nejjari, Mohammed</creatorcontrib><creatorcontrib>Pillière, Remy</creatorcontrib><creatorcontrib>Brochet, Eric</creatorcontrib><creatorcontrib>Mangin, Lionel</creatorcontrib><creatorcontrib>Bonnet, Guillaume</creatorcontrib><creatorcontrib>Ciobotaru, Vlad</creatorcontrib><creatorcontrib>Leurent, Guillaume</creatorcontrib><creatorcontrib>Hammoudi, Najib</creatorcontrib><creatorcontrib>Aminian, Adel</creatorcontrib><creatorcontrib>Karsenty, Clement</creatorcontrib><creatorcontrib>Spaulding, Christian</creatorcontrib><creatorcontrib>Armero, Sebastien</creatorcontrib><creatorcontrib>Collet, Frederic</creatorcontrib><creatorcontrib>Champagnac, Didier</creatorcontrib><creatorcontrib>Ternacle, Julien</creatorcontrib><creatorcontrib>Kloeckner, Martin</creatorcontrib><creatorcontrib>Gerardin, Benoit</creatorcontrib><creatorcontrib>Isorni, Marc-Antoine</creatorcontrib><title>Multimodality imaging guidance for percutaneous paravalvular leak closure: Insights from the multi-centre FFPP register</title><title>Archives of cardiovascular diseases</title><addtitle>Arch Cardiovasc Dis</addtitle><description>Percutaneous paravalvular leak (PVL) closure has emerged as a palliative alternative to surgical management in selected high-risk patients. Percutaneous procedures are challenging, especially for mitral PVL. Accurate imaging of the morphologies of the defects is mandatory, together with precise guidance in the catheterization laboratory to enhance success rates.
To describe imaging modalities used in clinical practice to guide percutaneous PVL closure and assess the potential of new imaging tools.
Data from the ‘Fermeture de Fuite paraprothétique’ (FFPP) register were used. The FFPP register is an international multi-institutional collaborative register started in 2017 with a retrospective and a prospective part. A descriptive analysis of multimodality imaging used to guide PVL closure in clinical practice was performed.
Data from 173 procedures performed in 19 centres from three countries (France, Belgium and Poland) were collected, which included eight cases of PVL following transcatheter valve replacement. Transoesophageal echocardiography was used in 167 cases (96.5%) and 3D echocardiography in 87.4% of cases. In one case, 3D-echocardiography was fused with fluoroscopy images in real time using echonavigator software. Details about multimodality imaging were available from a sample of 31 patients. Cardiac computed tomography (CT) was performed before 10 of the procedures. In one case, fusion between preprocedural cardiac CT angiography data and fluoroscopy data was used. In two cases, a 3D model of the valve with PVL was printed.
Echocardiography, particularly the 3D mode, is the cornerstone of PVL imaging. Other imaging modalities, such as cardiac CT and cardiac magnetic resonance imaging, may be of complementary interest. New techniques such as imaging fusion and printing may further facilitate the percutaneous approach of PVLs.
La fermeture percutanée des fuites valvulaires paraprothétiques (FPP) est devenue une alternative à la chirurgie chez des patients sélectionnés à haut-risque. Les procédures sont complexes en particulier pour les FPP mitrales. Une imagerie de précision est nécessaire pour décrire la morphologie des défects, guider la procédure percutanée et augmenter le taux de succès.
L’objectif de cette étude est de décrire les différents examens d’imagerie utilisés en pratique courante et leur intérêt respectifs ainsi que l’éventuel apport des nouvelles techniques.
Les données du registre « fermeture de fuite paraprothétique » (FFPP) ont été utilisées. Ce registre multicentrique international débuté en 2017 comporte un bras rétrospectif et un bras prospectif. Les examens d’imagerie utilisés pour la fermeture percutanée des FPP ont été décrits.
L’étude a inclus 173 procédures réalisées dans 19 centres de trois pays, incluant huit procédures sur des FPP sur valves implantées par voie percutanée. L’échographie transoesophagienne a été utilisée dans 167 cas (96,5 %) dont 87,4 % avec le mode 3D. Dans un cas, l’échocardiographie 3D a été fusionnée avec la fluoroscopie en temps réel. Au sein d’un échantillon de 31 patients dont les données des différents examens d’imagerie étaient disponibles, 10 scanners cardiaques ont été réalisés. Dans un cas, ce scanner a été fusionné avec la fluoroscopie durant la procédure. Deux modèles 3D de valves avec FPP ont été imprimés.
L’échocardiographie et en particulier le mode 3D est la pierre angulaire de l’imagerie des FPP. Les autres techniques d’imagerie incluant le scanner et l’IRM sont complémentaires. Les nouvelles techniques incluant la fusion et l’impression 3D contribueront possiblement à mieux appréhender les FPP et leur traitement percutané.</description><subject>3D echocardiography</subject><subject>4D-flow MRI</subject><subject>Cardiac valve</subject><subject>Echocardiographie 3D</subject><subject>Fuite paraprothétique</subject><subject>IRM 4D flow</subject><subject>Life Sciences</subject><subject>Paravalvular leak</subject><subject>Valve cardiaque</subject><issn>1875-2136</issn><issn>1875-2128</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><recordid>eNp9kcFu1DAURSMEoqXwAyyQl7BIsB3HiRGbqmLaSoPoAtaW47xkPDjxYDtB_XscpZ0lK1tP513pvpNl7wkuCCb887FQeukKiklT4KrAmLzILklTVzkltHl5_pf8InsTwhFjTuuav84uqBBlTXF9mf39PttoRtcpa-IjMqMazDSgYTadmjSg3nl0Aq_nqCZwc0An5dWi7DJb5ZEF9Rtp68Ls4Qu6n4IZDjGg3rsRxQOgcQ3PNUzRA9rtHh6Qh8GECP5t9qpXNsC7p_cq-7X79vPmLt__uL2_ud7nmpUs5k1JKWcl17pjmrGeC657qBpOO1G2lWK0FiBaRhTFTOlG9KxtgTVpzvpUt7zKPm25B2Xlyad-_lE6ZeTd9V6uM8xKUVHCFpLYjxt78u7PDCHK0QQN1m7VJcWVSDiv64TSDdXeheChP2cTLFc58ihXOXKVI3Elk5y09OEpf25H6M4rzzYS8HUDIF1kMeBl0AaShs540FF2zvwv_x9xaaD7</recordid><startdate>201806</startdate><enddate>201806</enddate><creator>Hascoet, Sébastien</creator><creator>Smolka, Grzegorz</creator><creator>Bagate, Francois</creator><creator>Guihaire, Julien</creator><creator>Potier, Agathe</creator><creator>Hadeed, Khaled</creator><creator>Lavie-Badie, Yoan</creator><creator>Bouvaist, Hélène</creator><creator>Dauphin, Claire</creator><creator>Bauer, Fabrice</creator><creator>Nejjari, Mohammed</creator><creator>Pillière, Remy</creator><creator>Brochet, Eric</creator><creator>Mangin, Lionel</creator><creator>Bonnet, Guillaume</creator><creator>Ciobotaru, Vlad</creator><creator>Leurent, Guillaume</creator><creator>Hammoudi, Najib</creator><creator>Aminian, Adel</creator><creator>Karsenty, Clement</creator><creator>Spaulding, Christian</creator><creator>Armero, Sebastien</creator><creator>Collet, Frederic</creator><creator>Champagnac, Didier</creator><creator>Ternacle, Julien</creator><creator>Kloeckner, Martin</creator><creator>Gerardin, Benoit</creator><creator>Isorni, Marc-Antoine</creator><general>Elsevier Masson SAS</general><general>Elsevier ; Société française de cardiologie [2008-....]</general><scope>6I.</scope><scope>AAFTH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>1XC</scope><orcidid>https://orcid.org/0000-0001-6506-1224</orcidid></search><sort><creationdate>201806</creationdate><title>Multimodality imaging guidance for percutaneous paravalvular leak closure: Insights from the multi-centre FFPP register</title><author>Hascoet, Sébastien ; Smolka, Grzegorz ; Bagate, Francois ; Guihaire, Julien ; Potier, Agathe ; Hadeed, Khaled ; Lavie-Badie, Yoan ; Bouvaist, Hélène ; Dauphin, Claire ; Bauer, Fabrice ; Nejjari, Mohammed ; Pillière, Remy ; Brochet, Eric ; Mangin, Lionel ; Bonnet, Guillaume ; Ciobotaru, Vlad ; Leurent, Guillaume ; Hammoudi, Najib ; Aminian, Adel ; Karsenty, Clement ; Spaulding, Christian ; Armero, Sebastien ; Collet, Frederic ; Champagnac, Didier ; Ternacle, Julien ; Kloeckner, Martin ; Gerardin, Benoit ; Isorni, Marc-Antoine</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c434t-83226436ccd4c44f696cfe5862d93b5a4279e9b41a204ac89f4bbe484274f0623</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>3D echocardiography</topic><topic>4D-flow MRI</topic><topic>Cardiac valve</topic><topic>Echocardiographie 3D</topic><topic>Fuite paraprothétique</topic><topic>IRM 4D flow</topic><topic>Life Sciences</topic><topic>Paravalvular leak</topic><topic>Valve cardiaque</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Hascoet, Sébastien</creatorcontrib><creatorcontrib>Smolka, Grzegorz</creatorcontrib><creatorcontrib>Bagate, Francois</creatorcontrib><creatorcontrib>Guihaire, Julien</creatorcontrib><creatorcontrib>Potier, Agathe</creatorcontrib><creatorcontrib>Hadeed, Khaled</creatorcontrib><creatorcontrib>Lavie-Badie, Yoan</creatorcontrib><creatorcontrib>Bouvaist, Hélène</creatorcontrib><creatorcontrib>Dauphin, Claire</creatorcontrib><creatorcontrib>Bauer, Fabrice</creatorcontrib><creatorcontrib>Nejjari, Mohammed</creatorcontrib><creatorcontrib>Pillière, Remy</creatorcontrib><creatorcontrib>Brochet, Eric</creatorcontrib><creatorcontrib>Mangin, Lionel</creatorcontrib><creatorcontrib>Bonnet, Guillaume</creatorcontrib><creatorcontrib>Ciobotaru, Vlad</creatorcontrib><creatorcontrib>Leurent, Guillaume</creatorcontrib><creatorcontrib>Hammoudi, Najib</creatorcontrib><creatorcontrib>Aminian, Adel</creatorcontrib><creatorcontrib>Karsenty, Clement</creatorcontrib><creatorcontrib>Spaulding, Christian</creatorcontrib><creatorcontrib>Armero, Sebastien</creatorcontrib><creatorcontrib>Collet, Frederic</creatorcontrib><creatorcontrib>Champagnac, Didier</creatorcontrib><creatorcontrib>Ternacle, Julien</creatorcontrib><creatorcontrib>Kloeckner, Martin</creatorcontrib><creatorcontrib>Gerardin, Benoit</creatorcontrib><creatorcontrib>Isorni, Marc-Antoine</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Hyper Article en Ligne (HAL)</collection><jtitle>Archives of cardiovascular diseases</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Hascoet, Sébastien</au><au>Smolka, Grzegorz</au><au>Bagate, Francois</au><au>Guihaire, Julien</au><au>Potier, Agathe</au><au>Hadeed, Khaled</au><au>Lavie-Badie, Yoan</au><au>Bouvaist, Hélène</au><au>Dauphin, Claire</au><au>Bauer, Fabrice</au><au>Nejjari, Mohammed</au><au>Pillière, Remy</au><au>Brochet, Eric</au><au>Mangin, Lionel</au><au>Bonnet, Guillaume</au><au>Ciobotaru, Vlad</au><au>Leurent, Guillaume</au><au>Hammoudi, Najib</au><au>Aminian, Adel</au><au>Karsenty, Clement</au><au>Spaulding, Christian</au><au>Armero, Sebastien</au><au>Collet, Frederic</au><au>Champagnac, Didier</au><au>Ternacle, Julien</au><au>Kloeckner, Martin</au><au>Gerardin, Benoit</au><au>Isorni, Marc-Antoine</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Multimodality imaging guidance for percutaneous paravalvular leak closure: Insights from the multi-centre FFPP register</atitle><jtitle>Archives of cardiovascular diseases</jtitle><addtitle>Arch Cardiovasc Dis</addtitle><date>2018-06</date><risdate>2018</risdate><volume>111</volume><issue>6-7</issue><spage>421</spage><epage>431</epage><pages>421-431</pages><issn>1875-2136</issn><eissn>1875-2128</eissn><abstract>Percutaneous paravalvular leak (PVL) closure has emerged as a palliative alternative to surgical management in selected high-risk patients. Percutaneous procedures are challenging, especially for mitral PVL. Accurate imaging of the morphologies of the defects is mandatory, together with precise guidance in the catheterization laboratory to enhance success rates.
To describe imaging modalities used in clinical practice to guide percutaneous PVL closure and assess the potential of new imaging tools.
Data from the ‘Fermeture de Fuite paraprothétique’ (FFPP) register were used. The FFPP register is an international multi-institutional collaborative register started in 2017 with a retrospective and a prospective part. A descriptive analysis of multimodality imaging used to guide PVL closure in clinical practice was performed.
Data from 173 procedures performed in 19 centres from three countries (France, Belgium and Poland) were collected, which included eight cases of PVL following transcatheter valve replacement. Transoesophageal echocardiography was used in 167 cases (96.5%) and 3D echocardiography in 87.4% of cases. In one case, 3D-echocardiography was fused with fluoroscopy images in real time using echonavigator software. Details about multimodality imaging were available from a sample of 31 patients. Cardiac computed tomography (CT) was performed before 10 of the procedures. In one case, fusion between preprocedural cardiac CT angiography data and fluoroscopy data was used. In two cases, a 3D model of the valve with PVL was printed.
Echocardiography, particularly the 3D mode, is the cornerstone of PVL imaging. Other imaging modalities, such as cardiac CT and cardiac magnetic resonance imaging, may be of complementary interest. New techniques such as imaging fusion and printing may further facilitate the percutaneous approach of PVLs.
La fermeture percutanée des fuites valvulaires paraprothétiques (FPP) est devenue une alternative à la chirurgie chez des patients sélectionnés à haut-risque. Les procédures sont complexes en particulier pour les FPP mitrales. Une imagerie de précision est nécessaire pour décrire la morphologie des défects, guider la procédure percutanée et augmenter le taux de succès.
L’objectif de cette étude est de décrire les différents examens d’imagerie utilisés en pratique courante et leur intérêt respectifs ainsi que l’éventuel apport des nouvelles techniques.
Les données du registre « fermeture de fuite paraprothétique » (FFPP) ont été utilisées. Ce registre multicentrique international débuté en 2017 comporte un bras rétrospectif et un bras prospectif. Les examens d’imagerie utilisés pour la fermeture percutanée des FPP ont été décrits.
L’étude a inclus 173 procédures réalisées dans 19 centres de trois pays, incluant huit procédures sur des FPP sur valves implantées par voie percutanée. L’échographie transoesophagienne a été utilisée dans 167 cas (96,5 %) dont 87,4 % avec le mode 3D. Dans un cas, l’échocardiographie 3D a été fusionnée avec la fluoroscopie en temps réel. Au sein d’un échantillon de 31 patients dont les données des différents examens d’imagerie étaient disponibles, 10 scanners cardiaques ont été réalisés. Dans un cas, ce scanner a été fusionné avec la fluoroscopie durant la procédure. Deux modèles 3D de valves avec FPP ont été imprimés.
L’échocardiographie et en particulier le mode 3D est la pierre angulaire de l’imagerie des FPP. Les autres techniques d’imagerie incluant le scanner et l’IRM sont complémentaires. Les nouvelles techniques incluant la fusion et l’impression 3D contribueront possiblement à mieux appréhender les FPP et leur traitement percutané.</abstract><cop>Netherlands</cop><pub>Elsevier Masson SAS</pub><pmid>29937207</pmid><doi>10.1016/j.acvd.2018.05.001</doi><tpages>11</tpages><orcidid>https://orcid.org/0000-0001-6506-1224</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1875-2136 |
ispartof | Archives of cardiovascular diseases, 2018-06, Vol.111 (6-7), p.421-431 |
issn | 1875-2136 1875-2128 |
language | eng |
recordid | cdi_hal_primary_oai_HAL_hal_04395214v1 |
source | ScienceDirect Freedom Collection |
subjects | 3D echocardiography 4D-flow MRI Cardiac valve Echocardiographie 3D Fuite paraprothétique IRM 4D flow Life Sciences Paravalvular leak Valve cardiaque |
title | Multimodality imaging guidance for percutaneous paravalvular leak closure: Insights from the multi-centre FFPP register |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T05%3A03%3A06IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_hal_p&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Multimodality%20imaging%20guidance%20for%20percutaneous%20paravalvular%20leak%20closure:%20Insights%20from%20the%20multi-centre%20FFPP%20register&rft.jtitle=Archives%20of%20cardiovascular%20diseases&rft.au=Hascoet,%20S%C3%A9bastien&rft.date=2018-06&rft.volume=111&rft.issue=6-7&rft.spage=421&rft.epage=431&rft.pages=421-431&rft.issn=1875-2136&rft.eissn=1875-2128&rft_id=info:doi/10.1016/j.acvd.2018.05.001&rft_dat=%3Cproquest_hal_p%3E2059043677%3C/proquest_hal_p%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c434t-83226436ccd4c44f696cfe5862d93b5a4279e9b41a204ac89f4bbe484274f0623%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2059043677&rft_id=info:pmid/29937207&rfr_iscdi=true |