Loading…
Differential Measurement System based on Programmable Josephson Voltage Standards for low frequency ac voltage measurement
The paper discusses the validation of ac voltage differential measurement system based on Programmable Josephson Voltage Standards and commercial ac sources. It focuses on calibrating the fundamental effective value of low-frequency (62.5 Hz and 100 Hz) ac sine waves (0.75 Vrms). Preliminary results...
Saved in:
Main Authors: | , , , , , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The paper discusses the validation of ac voltage differential measurement system based on Programmable Josephson Voltage Standards and commercial ac sources. It focuses on calibrating the fundamental effective value of low-frequency (62.5 Hz and 100 Hz) ac sine waves (0.75 Vrms). Preliminary results show that the relative consistency and Type A uncertainty under these conditions are better than 1 part in 10 6 . |
---|---|
ISSN: | 2160-0171 |
DOI: | 10.1109/CPEM61406.2024.10646115 |