Reducing the effects of scattered photons in X-ray projection imaging
An iterative algorithm is introduced which can compute and subtract the scatter bias. It is assumed that the scatter-free image is spatially bounded and anything outside the boundary is the result of recorded scattered photons. This information and the fact that the scatter distribution does not hav...
Saved in:
Published in: | IEEE transactions on medical imaging 1989-03, Vol.8 (1), p.56-63 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An iterative algorithm is introduced which can compute and subtract the scatter bias. It is assumed that the scatter-free image is spatially bounded and anything outside the boundary is the result of recorded scattered photons. This information and the fact that the scatter distribution does not have appreciable high-frequency content are utilized.< > |
---|---|
ISSN: | 0278-0062 1558-254X |
DOI: | 10.1109/42.20362 |