Intercomparison of Precision Current Sources in the Range From 10 fA to 10 pA Between PTB and VNIIM

An intercomparison of precision current sources was carried out for currents between 10 fA to 10 pA. The current sources of both PTB and VNIIM were based on the principle of charging a capacitor using a linear voltage ramp (Rozhdestvenskaya et al., 1979) and (Willenberg et al., 2003). The relative d...

Full description

Saved in:
Bibliographic Details
Main Authors: Willenberg, G.-D., Tauscher, H.N., Pavlov, O.M., Korotkova, I.
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An intercomparison of precision current sources was carried out for currents between 10 fA to 10 pA. The current sources of both PTB and VNIIM were based on the principle of charging a capacitor using a linear voltage ramp (Rozhdestvenskaya et al., 1979) and (Willenberg et al., 2003). The relative deviation for the values of the sources were found to be between 2.8middot10 -4 and 3.1middot10 -3 , depending on the current range