Uncertainty Evaluation of a Broadband Attenuation Standard
We have recently developed a standard step attenuator, composed of 10 steps of build-up chain of power measurement and operating from 10 MHz to 18 GHz. Each step shows a nominal attenuation of about 9 dB and an expanded uncertainty of 0.0018 dB at 30 MHz. The overall system has a dynamic range over...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We have recently developed a standard step attenuator, composed of 10 steps of build-up chain of power measurement and operating from 10 MHz to 18 GHz. Each step shows a nominal attenuation of about 9 dB and an expanded uncertainty of 0.0018 dB at 30 MHz. The overall system has a dynamic range over 90 dB |
---|---|