Fully implantable blood pressure sensor for hypertonic patients

In Germany approximately ten million people suffer from high blood pressure. About 1% of these patients can hardly be stabilized on drugs and a long time monitoring is advisable. A system consisting of an implant and an external reader station is presented to realize a long term monitoring for these...

Full description

Saved in:
Bibliographic Details
Main Authors: Fassbender, H., Mokwa, W., Gortz, M., Trieu, K., Urban, U., Schmitz-Rode, T., Gottsche, T., Osypka, P.
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page 1229
container_issue
container_start_page 1226
container_title
container_volume
creator Fassbender, H.
Mokwa, W.
Gortz, M.
Trieu, K.
Urban, U.
Schmitz-Rode, T.
Gottsche, T.
Osypka, P.
description In Germany approximately ten million people suffer from high blood pressure. About 1% of these patients can hardly be stabilized on drugs and a long time monitoring is advisable. A system consisting of an implant and an external reader station is presented to realize a long term monitoring for these patients. The implant consists of a newly designed sensor chip integrated at the head of a catheter (Oslash 1.1 mm) and a telemetric unit. The sensor-tip is placed into the femoral artery while the telemetric unit is implanted into the subcutaneous tissue. Thus the disturbance inside the blood vessel and the distance for wireless communication are kept as small as possible to obtain optimal parameters. The implant is supplied with energy wirelessly via inductive coupling from the external reader station. Data is readout from the external station with approx. 30 Hz and an overall accuracy of plusmn 2 mbar.
doi_str_mv 10.1109/ICSENS.2008.4716664
format conference_proceeding
fullrecord <record><control><sourceid>ieee_CHZPO</sourceid><recordid>TN_cdi_ieee_primary_4716664</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><ieee_id>4716664</ieee_id><sourcerecordid>4716664</sourcerecordid><originalsourceid>FETCH-LOGICAL-i90t-6e72cd4679ab45c2c4b0a24aed9caea0fd8d5bb8c9409b9a7c3759dd293cc60d3</originalsourceid><addsrcrecordid>eNpFkL1qwzAURtU_qJP2CbL4BexeybKsO5VikjYQ2iHZg36uqYtiG8sZ_PYtNJDh4wwHzvAxtuKQcw74sq336899LgB0LiuulJI3bMGlkFKUmqtblgiudIZC4N1VgL5nCccCMiiwfGSLGH8ABJRCJ-x1cw5hTtvTEEw3GRsotaHvfTqMFON5pDRSF_sxbf72PQ80Tn3XunQwU0vdFJ_YQ2NCpOcLl-ywWR_qj2z39b6t33ZZizBliirhvFQVGitLJ5y0YIQ05NEZMtB47UtrtUMJaNFUrqhK9F5g4ZwCXyzZ6j_bEtFxGNuTGefj5YTiF3pCTtM</addsrcrecordid><sourcetype>Publisher</sourcetype><iscdi>true</iscdi><recordtype>conference_proceeding</recordtype></control><display><type>conference_proceeding</type><title>Fully implantable blood pressure sensor for hypertonic patients</title><source>IEEE Xplore All Conference Series</source><creator>Fassbender, H. ; Mokwa, W. ; Gortz, M. ; Trieu, K. ; Urban, U. ; Schmitz-Rode, T. ; Gottsche, T. ; Osypka, P.</creator><creatorcontrib>Fassbender, H. ; Mokwa, W. ; Gortz, M. ; Trieu, K. ; Urban, U. ; Schmitz-Rode, T. ; Gottsche, T. ; Osypka, P.</creatorcontrib><description>In Germany approximately ten million people suffer from high blood pressure. About 1% of these patients can hardly be stabilized on drugs and a long time monitoring is advisable. A system consisting of an implant and an external reader station is presented to realize a long term monitoring for these patients. The implant consists of a newly designed sensor chip integrated at the head of a catheter (Oslash 1.1 mm) and a telemetric unit. The sensor-tip is placed into the femoral artery while the telemetric unit is implanted into the subcutaneous tissue. Thus the disturbance inside the blood vessel and the distance for wireless communication are kept as small as possible to obtain optimal parameters. The implant is supplied with energy wirelessly via inductive coupling from the external reader station. Data is readout from the external station with approx. 30 Hz and an overall accuracy of plusmn 2 mbar.</description><identifier>ISSN: 1930-0395</identifier><identifier>ISBN: 1424425808</identifier><identifier>ISBN: 9781424425808</identifier><identifier>EISSN: 2168-9229</identifier><identifier>EISBN: 1424425816</identifier><identifier>EISBN: 9781424425815</identifier><identifier>DOI: 10.1109/ICSENS.2008.4716664</identifier><language>eng</language><publisher>IEEE</publisher><subject>Arteries ; Biomedical monitoring ; Blood pressure ; Blood vessels ; Catheters ; Drugs ; Implants ; Patient monitoring ; Telemetry ; Wireless communication</subject><ispartof>2008 IEEE Sensors, 2008, p.1226-1229</ispartof><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://ieeexplore.ieee.org/document/4716664$$EHTML$$P50$$Gieee$$H</linktohtml><link.rule.ids>309,310,780,784,789,790,2058,27925,54555,54920,54932</link.rule.ids><linktorsrc>$$Uhttps://ieeexplore.ieee.org/document/4716664$$EView_record_in_IEEE$$FView_record_in_$$GIEEE</linktorsrc></links><search><creatorcontrib>Fassbender, H.</creatorcontrib><creatorcontrib>Mokwa, W.</creatorcontrib><creatorcontrib>Gortz, M.</creatorcontrib><creatorcontrib>Trieu, K.</creatorcontrib><creatorcontrib>Urban, U.</creatorcontrib><creatorcontrib>Schmitz-Rode, T.</creatorcontrib><creatorcontrib>Gottsche, T.</creatorcontrib><creatorcontrib>Osypka, P.</creatorcontrib><title>Fully implantable blood pressure sensor for hypertonic patients</title><title>2008 IEEE Sensors</title><addtitle>ICSENS</addtitle><description>In Germany approximately ten million people suffer from high blood pressure. About 1% of these patients can hardly be stabilized on drugs and a long time monitoring is advisable. A system consisting of an implant and an external reader station is presented to realize a long term monitoring for these patients. The implant consists of a newly designed sensor chip integrated at the head of a catheter (Oslash 1.1 mm) and a telemetric unit. The sensor-tip is placed into the femoral artery while the telemetric unit is implanted into the subcutaneous tissue. Thus the disturbance inside the blood vessel and the distance for wireless communication are kept as small as possible to obtain optimal parameters. The implant is supplied with energy wirelessly via inductive coupling from the external reader station. Data is readout from the external station with approx. 30 Hz and an overall accuracy of plusmn 2 mbar.</description><subject>Arteries</subject><subject>Biomedical monitoring</subject><subject>Blood pressure</subject><subject>Blood vessels</subject><subject>Catheters</subject><subject>Drugs</subject><subject>Implants</subject><subject>Patient monitoring</subject><subject>Telemetry</subject><subject>Wireless communication</subject><issn>1930-0395</issn><issn>2168-9229</issn><isbn>1424425808</isbn><isbn>9781424425808</isbn><isbn>1424425816</isbn><isbn>9781424425815</isbn><fulltext>true</fulltext><rsrctype>conference_proceeding</rsrctype><creationdate>2008</creationdate><recordtype>conference_proceeding</recordtype><sourceid>6IE</sourceid><recordid>eNpFkL1qwzAURtU_qJP2CbL4BexeybKsO5VikjYQ2iHZg36uqYtiG8sZ_PYtNJDh4wwHzvAxtuKQcw74sq336899LgB0LiuulJI3bMGlkFKUmqtblgiudIZC4N1VgL5nCccCMiiwfGSLGH8ABJRCJ-x1cw5hTtvTEEw3GRsotaHvfTqMFON5pDRSF_sxbf72PQ80Tn3XunQwU0vdFJ_YQ2NCpOcLl-ywWR_qj2z39b6t33ZZizBliirhvFQVGitLJ5y0YIQ05NEZMtB47UtrtUMJaNFUrqhK9F5g4ZwCXyzZ6j_bEtFxGNuTGefj5YTiF3pCTtM</recordid><startdate>200810</startdate><enddate>200810</enddate><creator>Fassbender, H.</creator><creator>Mokwa, W.</creator><creator>Gortz, M.</creator><creator>Trieu, K.</creator><creator>Urban, U.</creator><creator>Schmitz-Rode, T.</creator><creator>Gottsche, T.</creator><creator>Osypka, P.</creator><general>IEEE</general><scope>6IE</scope><scope>6IH</scope><scope>CBEJK</scope><scope>RIE</scope><scope>RIO</scope></search><sort><creationdate>200810</creationdate><title>Fully implantable blood pressure sensor for hypertonic patients</title><author>Fassbender, H. ; Mokwa, W. ; Gortz, M. ; Trieu, K. ; Urban, U. ; Schmitz-Rode, T. ; Gottsche, T. ; Osypka, P.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-i90t-6e72cd4679ab45c2c4b0a24aed9caea0fd8d5bb8c9409b9a7c3759dd293cc60d3</frbrgroupid><rsrctype>conference_proceedings</rsrctype><prefilter>conference_proceedings</prefilter><language>eng</language><creationdate>2008</creationdate><topic>Arteries</topic><topic>Biomedical monitoring</topic><topic>Blood pressure</topic><topic>Blood vessels</topic><topic>Catheters</topic><topic>Drugs</topic><topic>Implants</topic><topic>Patient monitoring</topic><topic>Telemetry</topic><topic>Wireless communication</topic><toplevel>online_resources</toplevel><creatorcontrib>Fassbender, H.</creatorcontrib><creatorcontrib>Mokwa, W.</creatorcontrib><creatorcontrib>Gortz, M.</creatorcontrib><creatorcontrib>Trieu, K.</creatorcontrib><creatorcontrib>Urban, U.</creatorcontrib><creatorcontrib>Schmitz-Rode, T.</creatorcontrib><creatorcontrib>Gottsche, T.</creatorcontrib><creatorcontrib>Osypka, P.</creatorcontrib><collection>IEEE Electronic Library (IEL) Conference Proceedings</collection><collection>IEEE Proceedings Order Plan (POP) 1998-present by volume</collection><collection>IEEE Xplore All Conference Proceedings</collection><collection>IEEE Electronic Library Online</collection><collection>IEEE Proceedings Order Plans (POP) 1998-present</collection></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext_linktorsrc</fulltext></delivery><addata><au>Fassbender, H.</au><au>Mokwa, W.</au><au>Gortz, M.</au><au>Trieu, K.</au><au>Urban, U.</au><au>Schmitz-Rode, T.</au><au>Gottsche, T.</au><au>Osypka, P.</au><format>book</format><genre>proceeding</genre><ristype>CONF</ristype><atitle>Fully implantable blood pressure sensor for hypertonic patients</atitle><btitle>2008 IEEE Sensors</btitle><stitle>ICSENS</stitle><date>2008-10</date><risdate>2008</risdate><spage>1226</spage><epage>1229</epage><pages>1226-1229</pages><issn>1930-0395</issn><eissn>2168-9229</eissn><isbn>1424425808</isbn><isbn>9781424425808</isbn><eisbn>1424425816</eisbn><eisbn>9781424425815</eisbn><abstract>In Germany approximately ten million people suffer from high blood pressure. About 1% of these patients can hardly be stabilized on drugs and a long time monitoring is advisable. A system consisting of an implant and an external reader station is presented to realize a long term monitoring for these patients. The implant consists of a newly designed sensor chip integrated at the head of a catheter (Oslash 1.1 mm) and a telemetric unit. The sensor-tip is placed into the femoral artery while the telemetric unit is implanted into the subcutaneous tissue. Thus the disturbance inside the blood vessel and the distance for wireless communication are kept as small as possible to obtain optimal parameters. The implant is supplied with energy wirelessly via inductive coupling from the external reader station. Data is readout from the external station with approx. 30 Hz and an overall accuracy of plusmn 2 mbar.</abstract><pub>IEEE</pub><doi>10.1109/ICSENS.2008.4716664</doi><tpages>4</tpages></addata></record>
fulltext fulltext_linktorsrc
identifier ISSN: 1930-0395
ispartof 2008 IEEE Sensors, 2008, p.1226-1229
issn 1930-0395
2168-9229
language eng
recordid cdi_ieee_primary_4716664
source IEEE Xplore All Conference Series
subjects Arteries
Biomedical monitoring
Blood pressure
Blood vessels
Catheters
Drugs
Implants
Patient monitoring
Telemetry
Wireless communication
title Fully implantable blood pressure sensor for hypertonic patients
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T22%3A26%3A04IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-ieee_CHZPO&rft_val_fmt=info:ofi/fmt:kev:mtx:book&rft.genre=proceeding&rft.atitle=Fully%20implantable%20blood%20pressure%20sensor%20for%20hypertonic%20patients&rft.btitle=2008%20IEEE%20Sensors&rft.au=Fassbender,%20H.&rft.date=2008-10&rft.spage=1226&rft.epage=1229&rft.pages=1226-1229&rft.issn=1930-0395&rft.eissn=2168-9229&rft.isbn=1424425808&rft.isbn_list=9781424425808&rft_id=info:doi/10.1109/ICSENS.2008.4716664&rft.eisbn=1424425816&rft.eisbn_list=9781424425815&rft_dat=%3Cieee_CHZPO%3E4716664%3C/ieee_CHZPO%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-i90t-6e72cd4679ab45c2c4b0a24aed9caea0fd8d5bb8c9409b9a7c3759dd293cc60d3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rft_ieee_id=4716664&rfr_iscdi=true