Jaw Tracking System Using Resonated Wireless Markers
We have developed a jaw tracking system with wireless resonated markers. The system consists of two resonated markers, driving coil, 40 pickup coils, DA converter, and AD converters. The two resonated markers were composed of a coil with NiZn ferrite and tip condenser. We attached to one marker on t...
Saved in:
Published in: | IEEE transactions on magnetics 2009-10, Vol.45 (10), p.4880-4883 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We have developed a jaw tracking system with wireless resonated markers. The system consists of two resonated markers, driving coil, 40 pickup coils, DA converter, and AD converters. The two resonated markers were composed of a coil with NiZn ferrite and tip condenser. We attached to one marker on the incisor and the other on the forehead. Position accuracy of within 1 mm was achieved when the fabricated jaw model was moved about 40 mm. This system is available for tracking of reasonable jaw movement of normal subjects without disturbing the physiology. |
---|---|
ISSN: | 0018-9464 1941-0069 |
DOI: | 10.1109/TMAG.2009.2024637 |