A new permeance meter based on both lumped elements/transmission line theories

We have developed a new permeance meter in the 1 MHz-1.5 GHz range using either a microstrip pickup coil or a shielded loop pickup coil. Both pickup coils and the driving plates are free from LC resonance due to impedance matching. In the MHz range we must consider signal-to-noise ratio of the picku...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on magnetics 1996-09, Vol.32 (5), p.4941-4943
Main Authors: Yamaguchi, M., Yabukami, S., Arai, K.I.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We have developed a new permeance meter in the 1 MHz-1.5 GHz range using either a microstrip pickup coil or a shielded loop pickup coil. Both pickup coils and the driving plates are free from LC resonance due to impedance matching. In the MHz range we must consider signal-to-noise ratio of the pickup coil. In the GHz range we must eliminate the voltage induced by the electric field, and an electromagnetic higher order mode in the driving plates.
ISSN:0018-9464