Verification of performance of commercial LCR meters
Four high-accuracy LCR meters have been compared with the use of Metrosert and MIKES impedance standards. Specifications of LCR meters have been verified at frequency of 1 kHz for the resistance range from 1 Ω to 100 kΩ, for the capacitances in the range from 100 pF to 1 μF and for 100 mH inductor....
Saved in:
Main Authors: | , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Four high-accuracy LCR meters have been compared with the use of Metrosert and MIKES impedance standards. Specifications of LCR meters have been verified at frequency of 1 kHz for the resistance range from 1 Ω to 100 kΩ, for the capacitances in the range from 100 pF to 1 μF and for 100 mH inductor. The performance of LCR meters have been compared also at frequencies from 1 kHz to 1 MHz with ac resistors of different types, at 1 kHz and 100 kHz with 100 pF - 1 μF capacitance standards, and at 1 kHz with 300 μH - 300 mH inductance standards. Results of measurements presented in this summary confirm the specified uncertainty of the tested LCR meters. |
---|---|
ISSN: | 0589-1485 2160-0171 |
DOI: | 10.1109/CPEM.2010.5545242 |