Loading…

Design and accuracy evaluation of a new navigated drill system for computer assisted ENT-surgery

In this article a new navigated drill system for computer assisted ear, nose and throat (ENT) surgery is presented. The navigated drill and the microscope probe are part of a surgical navigation system for ENT-surgery. In particular, the accuracy of the new navigated drill is compared to an existing...

Full description

Saved in:
Bibliographic Details
Main Authors: Wenger, T., Nowatschin, S., Wittmann, W., Hurka, F., Strauss, G., Lueth, T. C.
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page 1236
container_issue
container_start_page 1233
container_title
container_volume 2011
creator Wenger, T.
Nowatschin, S.
Wittmann, W.
Hurka, F.
Strauss, G.
Lueth, T. C.
description In this article a new navigated drill system for computer assisted ear, nose and throat (ENT) surgery is presented. The navigated drill and the microscope probe are part of a surgical navigation system for ENT-surgery. In particular, the accuracy of the new navigated drill is compared to an existing navigated drill experimentally under conditions close to the surgical workflow. For the technical accuracy experiment, the new navigated drill in combination with the new microscope probe and a particular navigated measurement board have been integrated, together with the current navigated drill, in a navigation system by a special navigation software with measuring function, based on a standard ENT navigation software. The developed navigated measurement board provided the implementation of reproducible experiments and the direct accuracy comparison of the two navigated instruments under the same conditions. Thereby, N = 15 accuracy experiments are performed with both navigated drill systems with three possible tracker positions. The distance between the planned and the touched points were calculated and compared. The average distances from the planned points to the touched points with the new navigated drill is in the left tracker position 1.10 mm, in the middle tracker position 1.14 mm and in the right tracker position 1.59 mm. In comparison to the existing drill, the new navigated drill, measured with each tracker position, is 0.62 mm more accurate.
doi_str_mv 10.1109/IEMBS.2011.6090290
format conference_proceeding
fullrecord <record><control><sourceid>proquest_6IE</sourceid><recordid>TN_cdi_ieee_primary_6090290</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><ieee_id>6090290</ieee_id><sourcerecordid>916855145</sourcerecordid><originalsourceid>FETCH-LOGICAL-i300t-e79d56941e033adbfc1ddd32cc61f53f791f4b21e4b69608d83261bf4765b2d53</originalsourceid><addsrcrecordid>eNo1kDtPwzAUhc1LUEr_ACzemFJ8_Uo88ihQqcBAkdiKE99URmlS7KSo_55WLWc5w_fpDIeQS2BDAGZuxqOXu_chZwBDzQzjhh2Qc5BcSgmc60PSA6WyRGpQR2Rg0uyfARxvGDMy0Vn6eUYGMX6zTbQ2QvBTcsY5V1IJ0yNfDxj9vKa2dtQWRRdssaa4slVnW9_UtCmppTX-0tqu_Ny26KgLvqpoXMcWF7RsAi2axbJrMVAbo49bZfQ6TWIX5hjWF-SktFXEwb775ONxNL1_TiZvT-P720niBWNtgqlxShsJyISwLi8LcM4JXhQaSiXK1EApcw4oc200y1wmuIa8lKlWOXdK9Mn1bncZmp8OYztb-FhgVdkamy7ODOhMKZBb82pnekScLYNf2LCe7R8Wf0wuaX8</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>conference_proceeding</recordtype><pqid>916855145</pqid></control><display><type>conference_proceeding</type><title>Design and accuracy evaluation of a new navigated drill system for computer assisted ENT-surgery</title><source>IEEE Electronic Library (IEL) Conference Proceedings</source><creator>Wenger, T. ; Nowatschin, S. ; Wittmann, W. ; Hurka, F. ; Strauss, G. ; Lueth, T. C.</creator><creatorcontrib>Wenger, T. ; Nowatschin, S. ; Wittmann, W. ; Hurka, F. ; Strauss, G. ; Lueth, T. C.</creatorcontrib><description>In this article a new navigated drill system for computer assisted ear, nose and throat (ENT) surgery is presented. The navigated drill and the microscope probe are part of a surgical navigation system for ENT-surgery. In particular, the accuracy of the new navigated drill is compared to an existing navigated drill experimentally under conditions close to the surgical workflow. For the technical accuracy experiment, the new navigated drill in combination with the new microscope probe and a particular navigated measurement board have been integrated, together with the current navigated drill, in a navigation system by a special navigation software with measuring function, based on a standard ENT navigation software. The developed navigated measurement board provided the implementation of reproducible experiments and the direct accuracy comparison of the two navigated instruments under the same conditions. Thereby, N = 15 accuracy experiments are performed with both navigated drill systems with three possible tracker positions. The distance between the planned and the touched points were calculated and compared. The average distances from the planned points to the touched points with the new navigated drill is in the left tracker position 1.10 mm, in the middle tracker position 1.14 mm and in the right tracker position 1.59 mm. In comparison to the existing drill, the new navigated drill, measured with each tracker position, is 0.62 mm more accurate.</description><identifier>ISSN: 1094-687X</identifier><identifier>ISBN: 9781424441211</identifier><identifier>ISBN: 1424441218</identifier><identifier>EISSN: 1558-4615</identifier><identifier>EISSN: 2694-0604</identifier><identifier>EISBN: 1424441226</identifier><identifier>EISBN: 1457715899</identifier><identifier>EISBN: 9781457715891</identifier><identifier>EISBN: 9781424441228</identifier><identifier>DOI: 10.1109/IEMBS.2011.6090290</identifier><identifier>PMID: 22254539</identifier><language>eng</language><publisher>IEEE</publisher><subject>Accuracy ; Instruments ; Navigation ; Optical variables measurement ; Position measurement ; Software measurement ; Surgery</subject><ispartof>2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2011, Vol.2011, p.1233-1236</ispartof><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://ieeexplore.ieee.org/document/6090290$$EHTML$$P50$$Gieee$$H</linktohtml><link.rule.ids>309,310,314,780,784,789,790,2057,27923,27924,54919</link.rule.ids><linktorsrc>$$Uhttps://ieeexplore.ieee.org/document/6090290$$EView_record_in_IEEE$$FView_record_in_$$GIEEE</linktorsrc></links><search><creatorcontrib>Wenger, T.</creatorcontrib><creatorcontrib>Nowatschin, S.</creatorcontrib><creatorcontrib>Wittmann, W.</creatorcontrib><creatorcontrib>Hurka, F.</creatorcontrib><creatorcontrib>Strauss, G.</creatorcontrib><creatorcontrib>Lueth, T. C.</creatorcontrib><title>Design and accuracy evaluation of a new navigated drill system for computer assisted ENT-surgery</title><title>2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society</title><addtitle>IEMBS</addtitle><description>In this article a new navigated drill system for computer assisted ear, nose and throat (ENT) surgery is presented. The navigated drill and the microscope probe are part of a surgical navigation system for ENT-surgery. In particular, the accuracy of the new navigated drill is compared to an existing navigated drill experimentally under conditions close to the surgical workflow. For the technical accuracy experiment, the new navigated drill in combination with the new microscope probe and a particular navigated measurement board have been integrated, together with the current navigated drill, in a navigation system by a special navigation software with measuring function, based on a standard ENT navigation software. The developed navigated measurement board provided the implementation of reproducible experiments and the direct accuracy comparison of the two navigated instruments under the same conditions. Thereby, N = 15 accuracy experiments are performed with both navigated drill systems with three possible tracker positions. The distance between the planned and the touched points were calculated and compared. The average distances from the planned points to the touched points with the new navigated drill is in the left tracker position 1.10 mm, in the middle tracker position 1.14 mm and in the right tracker position 1.59 mm. In comparison to the existing drill, the new navigated drill, measured with each tracker position, is 0.62 mm more accurate.</description><subject>Accuracy</subject><subject>Instruments</subject><subject>Navigation</subject><subject>Optical variables measurement</subject><subject>Position measurement</subject><subject>Software measurement</subject><subject>Surgery</subject><issn>1094-687X</issn><issn>1558-4615</issn><issn>2694-0604</issn><isbn>9781424441211</isbn><isbn>1424441218</isbn><isbn>1424441226</isbn><isbn>1457715899</isbn><isbn>9781457715891</isbn><isbn>9781424441228</isbn><fulltext>true</fulltext><rsrctype>conference_proceeding</rsrctype><creationdate>2011</creationdate><recordtype>conference_proceeding</recordtype><sourceid>6IE</sourceid><recordid>eNo1kDtPwzAUhc1LUEr_ACzemFJ8_Uo88ihQqcBAkdiKE99URmlS7KSo_55WLWc5w_fpDIeQS2BDAGZuxqOXu_chZwBDzQzjhh2Qc5BcSgmc60PSA6WyRGpQR2Rg0uyfARxvGDMy0Vn6eUYGMX6zTbQ2QvBTcsY5V1IJ0yNfDxj9vKa2dtQWRRdssaa4slVnW9_UtCmppTX-0tqu_Ny26KgLvqpoXMcWF7RsAi2axbJrMVAbo49bZfQ6TWIX5hjWF-SktFXEwb775ONxNL1_TiZvT-P720niBWNtgqlxShsJyISwLi8LcM4JXhQaSiXK1EApcw4oc200y1wmuIa8lKlWOXdK9Mn1bncZmp8OYztb-FhgVdkamy7ODOhMKZBb82pnekScLYNf2LCe7R8Wf0wuaX8</recordid><startdate>20110101</startdate><enddate>20110101</enddate><creator>Wenger, T.</creator><creator>Nowatschin, S.</creator><creator>Wittmann, W.</creator><creator>Hurka, F.</creator><creator>Strauss, G.</creator><creator>Lueth, T. C.</creator><general>IEEE</general><scope>6IE</scope><scope>6IH</scope><scope>CBEJK</scope><scope>RIE</scope><scope>RIO</scope><scope>7X8</scope></search><sort><creationdate>20110101</creationdate><title>Design and accuracy evaluation of a new navigated drill system for computer assisted ENT-surgery</title><author>Wenger, T. ; Nowatschin, S. ; Wittmann, W. ; Hurka, F. ; Strauss, G. ; Lueth, T. C.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-i300t-e79d56941e033adbfc1ddd32cc61f53f791f4b21e4b69608d83261bf4765b2d53</frbrgroupid><rsrctype>conference_proceedings</rsrctype><prefilter>conference_proceedings</prefilter><language>eng</language><creationdate>2011</creationdate><topic>Accuracy</topic><topic>Instruments</topic><topic>Navigation</topic><topic>Optical variables measurement</topic><topic>Position measurement</topic><topic>Software measurement</topic><topic>Surgery</topic><toplevel>online_resources</toplevel><creatorcontrib>Wenger, T.</creatorcontrib><creatorcontrib>Nowatschin, S.</creatorcontrib><creatorcontrib>Wittmann, W.</creatorcontrib><creatorcontrib>Hurka, F.</creatorcontrib><creatorcontrib>Strauss, G.</creatorcontrib><creatorcontrib>Lueth, T. C.</creatorcontrib><collection>IEEE Electronic Library (IEL) Conference Proceedings</collection><collection>IEEE Proceedings Order Plan (POP) 1998-present by volume</collection><collection>IEEE Xplore All Conference Proceedings</collection><collection>IEEE Xplore</collection><collection>IEEE Proceedings Order Plans (POP) 1998-present</collection><collection>MEDLINE - Academic</collection></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext_linktorsrc</fulltext></delivery><addata><au>Wenger, T.</au><au>Nowatschin, S.</au><au>Wittmann, W.</au><au>Hurka, F.</au><au>Strauss, G.</au><au>Lueth, T. C.</au><format>book</format><genre>proceeding</genre><ristype>CONF</ristype><atitle>Design and accuracy evaluation of a new navigated drill system for computer assisted ENT-surgery</atitle><btitle>2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society</btitle><stitle>IEMBS</stitle><date>2011-01-01</date><risdate>2011</risdate><volume>2011</volume><spage>1233</spage><epage>1236</epage><pages>1233-1236</pages><issn>1094-687X</issn><eissn>1558-4615</eissn><eissn>2694-0604</eissn><isbn>9781424441211</isbn><isbn>1424441218</isbn><eisbn>1424441226</eisbn><eisbn>1457715899</eisbn><eisbn>9781457715891</eisbn><eisbn>9781424441228</eisbn><abstract>In this article a new navigated drill system for computer assisted ear, nose and throat (ENT) surgery is presented. The navigated drill and the microscope probe are part of a surgical navigation system for ENT-surgery. In particular, the accuracy of the new navigated drill is compared to an existing navigated drill experimentally under conditions close to the surgical workflow. For the technical accuracy experiment, the new navigated drill in combination with the new microscope probe and a particular navigated measurement board have been integrated, together with the current navigated drill, in a navigation system by a special navigation software with measuring function, based on a standard ENT navigation software. The developed navigated measurement board provided the implementation of reproducible experiments and the direct accuracy comparison of the two navigated instruments under the same conditions. Thereby, N = 15 accuracy experiments are performed with both navigated drill systems with three possible tracker positions. The distance between the planned and the touched points were calculated and compared. The average distances from the planned points to the touched points with the new navigated drill is in the left tracker position 1.10 mm, in the middle tracker position 1.14 mm and in the right tracker position 1.59 mm. In comparison to the existing drill, the new navigated drill, measured with each tracker position, is 0.62 mm more accurate.</abstract><pub>IEEE</pub><pmid>22254539</pmid><doi>10.1109/IEMBS.2011.6090290</doi><tpages>4</tpages></addata></record>
fulltext fulltext_linktorsrc
identifier ISSN: 1094-687X
ispartof 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2011, Vol.2011, p.1233-1236
issn 1094-687X
1558-4615
2694-0604
language eng
recordid cdi_ieee_primary_6090290
source IEEE Electronic Library (IEL) Conference Proceedings
subjects Accuracy
Instruments
Navigation
Optical variables measurement
Position measurement
Software measurement
Surgery
title Design and accuracy evaluation of a new navigated drill system for computer assisted ENT-surgery
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-08T09%3A12%3A10IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_6IE&rft_val_fmt=info:ofi/fmt:kev:mtx:book&rft.genre=proceeding&rft.atitle=Design%20and%20accuracy%20evaluation%20of%20a%20new%20navigated%20drill%20system%20for%20computer%20assisted%20ENT-surgery&rft.btitle=2011%20Annual%20International%20Conference%20of%20the%20IEEE%20Engineering%20in%20Medicine%20and%20Biology%20Society&rft.au=Wenger,%20T.&rft.date=2011-01-01&rft.volume=2011&rft.spage=1233&rft.epage=1236&rft.pages=1233-1236&rft.issn=1094-687X&rft.eissn=1558-4615&rft.isbn=9781424441211&rft.isbn_list=1424441218&rft_id=info:doi/10.1109/IEMBS.2011.6090290&rft.eisbn=1424441226&rft.eisbn_list=1457715899&rft.eisbn_list=9781457715891&rft.eisbn_list=9781424441228&rft_dat=%3Cproquest_6IE%3E916855145%3C/proquest_6IE%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-i300t-e79d56941e033adbfc1ddd32cc61f53f791f4b21e4b69608d83261bf4765b2d53%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=916855145&rft_id=info:pmid/22254539&rft_ieee_id=6090290&rfr_iscdi=true