THz absolute power measurement: A simple and reliable method
Achieving absolute power measurement is not so easy in the THz range. In this paper, we describe a simple method to perform power measurement no more expressed in a.u. but in Watt. For a 3.8 THz single-plasmon Quantum Cascade Laser, 28.8 mW peak power was measured with 12:1% relative uncertainty. Th...
Saved in:
Main Authors: | , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Achieving absolute power measurement is not so easy in the THz range. In this paper, we describe a simple method to perform power measurement no more expressed in a.u. but in Watt. For a 3.8 THz single-plasmon Quantum Cascade Laser, 28.8 mW peak power was measured with 12:1% relative uncertainty. The operating method described is not really a metrologic one but one that can be easily performed in every laboratory. |
---|---|
ISSN: | 2162-2027 |
DOI: | 10.1109/IRMMW-THz.2012.6380154 |