Loading…

Wireless Medical Telemetry Characterization for Ingestible Capsule Antenna Designs

This letter presents a comprehensive and systematic antenna performance evaluation and wireless medical telemetry characterization between two different ingestible antenna designs, an inverted conical helical antenna and a reference conformal meandered offset dipole antenna. The antenna performance...

Full description

Saved in:
Bibliographic Details
Published in:IEEE antennas and wireless propagation letters 2012, Vol.11, p.1679-1682
Main Authors: Rajagopalan, H., Rahmat-Samii, Y.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This letter presents a comprehensive and systematic antenna performance evaluation and wireless medical telemetry characterization between two different ingestible antenna designs, an inverted conical helical antenna and a reference conformal meandered offset dipole antenna. The antenna performance comparison matrix includes matching, radiation patterns, and polarization. Both these antennas are independently characterized inside a human body environment (box-model human). Specific absorption rate (SAR) evaluations are described, and finally a link budget is developed for both the antenna systems based on certain parameters (frequency, power limitations) with the goal of establishing a reliable communication link for indoor communication for biotelemetry applications.
ISSN:1536-1225
1548-5757
DOI:10.1109/LAWP.2013.2238502