Loading…

Optoacoustic determination of spatio- temporal responses of ultrasound sensors

The characterization of the spatial and frequency response of acoustic detectors is important for enabling accurate optoacoustic imaging. In this work, we developed a hybrid method for the characterization of the spatially dependent response of ultrasound detectors. The method is based on the experi...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control ferroelectrics, and frequency control, 2013-06, Vol.60 (6), p.1234-1244
Main Authors: Caballero, M. A. A., Rosenthal, A., Buehler, A., Razansky, D., Ntziachristos, V.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The characterization of the spatial and frequency response of acoustic detectors is important for enabling accurate optoacoustic imaging. In this work, we developed a hybrid method for the characterization of the spatially dependent response of ultrasound detectors. The method is based on the experimental determination of the receive-mode electrical impulse response (EIR) of the sensor, which is subsequently convolved with the corresponding spatial impulse response (SIR), computed numerically. The hybrid method is shown to have superior performance over purely experimental techniques in terms of accurate determination of the spatial and temporal responses of ultrasonic detectors, in high as well as low sensitivity regions of the sensor.
ISSN:0885-3010
1525-8955
DOI:10.1109/TUFFC.2013.2687