ECG noise reduction by ordered averaging

The authors propose a new method for signal averaging intended for nonstationary noise reduction in the ECG. The different ECG cycles are arranged in ascending order or noise variance before averaging. Then if only low-variance cycles are averaged, the improvement in signal-to-noise ratio is better...

Full description

Saved in:
Bibliographic Details
Main Authors: Ramos, J., Pallas-Areny, Y.
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The authors propose a new method for signal averaging intended for nonstationary noise reduction in the ECG. The different ECG cycles are arranged in ascending order or noise variance before averaging. Then if only low-variance cycles are averaged, the improvement in signal-to-noise ratio is better than averaging all cycles recorded.