Loading…
ECG noise reduction by ordered averaging
The authors propose a new method for signal averaging intended for nonstationary noise reduction in the ECG. The different ECG cycles are arranged in ascending order or noise variance before averaging. Then if only low-variance cycles are averaged, the improvement in signal-to-noise ratio is better...
Saved in:
Main Authors: | , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The authors propose a new method for signal averaging intended for nonstationary noise reduction in the ECG. The different ECG cycles are arranged in ascending order or noise variance before averaging. Then if only low-variance cycles are averaged, the improvement in signal-to-noise ratio is better than averaging all cycles recorded. |
---|---|
DOI: | 10.1109/IEMBS.1996.652660 |