Loading…
Intercomparison of a 10 mH inductance standard between UME-PTB
Turkish and German Metrology Institutes (UME and PTB) performed an intercomparison of 10 mH inductance standards at 1 kHz. The measurements of the thermostatically-controlled transfer standard of PTB were taken by new Maxwell-Wien bridge with a Wagner arm made in UME. Relative difference of UME and...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Turkish and German Metrology Institutes (UME and PTB) performed an intercomparison of 10 mH inductance standards at 1 kHz. The measurements of the thermostatically-controlled transfer standard of PTB were taken by new Maxwell-Wien bridge with a Wagner arm made in UME. Relative difference of UME and PTB measurements on the same standard was found to be 5.10/sup -6/. Results have shown that this type of Maxwell-Wien bridge is capable of producing very precise measurements on standard inductors and promised further developments that will increase the accuracy of the system. |
---|---|