Loading…
Automated comparator for AC-DC difference accurate measurements used at the BNM-LCIE
This paper describes the new automated comparator used at the BNM-LCIE for the calibration of voltage thermal converters with the best level of uncertainty. It allows to achieve A-type uncertainties of the order of a few parts in 10/sup 7/ in the frequency range from 10 Hz to 1 MHz (100 kHz for volt...
Saved in:
Main Author: | |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This paper describes the new automated comparator used at the BNM-LCIE for the calibration of voltage thermal converters with the best level of uncertainty. It allows to achieve A-type uncertainties of the order of a few parts in 10/sup 7/ in the frequency range from 10 Hz to 1 MHz (100 kHz for voltages above 100 V). B-type measurement uncertainty depends on the type of converters to be compared and on the applied voltage level, but generally is between a few parts in 10/sup 7/ and a few parts in 10/sup 6/. |
---|---|
DOI: | 10.1109/CPEM.1998.700031 |