Apparent backscatter of cancellous bone from fundamental and second harmonics using an ultrasound imaging system

This study aims to investigate the apparent backscatter of cancellous bone from the second ultrasonic harmonics using an imaging system. A cylinder model was used as the bone trabecular model and a positive correlation was found between the power spectrum of the second harmonics and trabecular thick...

Full description

Saved in:
Bibliographic Details
Main Authors: Yang Zheng, Xian Long, Jin, Craig, Nanan, Ralph, McEwan, Alistair
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This study aims to investigate the apparent backscatter of cancellous bone from the second ultrasonic harmonics using an imaging system. A cylinder model was used as the bone trabecular model and a positive correlation was found between the power spectrum of the second harmonics and trabecular thickness. The result of the in vitro experiment using ultrasound imaging system shows that the apparent backscatter parameters from the second harmonics (R=0.53- 0.71) correlates better with bone density than the fundamental backscatter signal (R=0.39-0.47).