A standard calibration system for geomagnetometers

A precise system (standard) for reproducing a DC magnetic field of the extended geomagnetic range (10-110 /spl mu/T) is established. The uncertainty of the total magnetic field reproduction in SI units is 3/spl times/10/sup -7/ inside a sphere 10 cm in diameter with an rms deviation of 0.002 nT. The...

Full description

Saved in:
Bibliographic Details
Main Authors: Shifrin, V.Ya, Alexandrov, E.B., Chikvadze, T.I., Kalabin, V.N., Yakobson, N.N., Khorev, V.N., Park, P.G.
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A precise system (standard) for reproducing a DC magnetic field of the extended geomagnetic range (10-110 /spl mu/T) is established. The uncertainty of the total magnetic field reproduction in SI units is 3/spl times/10/sup -7/ inside a sphere 10 cm in diameter with an rms deviation of 0.002 nT. The uncertainty of the magnetic field direction is 1 second of arc.
DOI:10.1109/CPEM.2000.850963