A Wireless Gauge Pressure Sensor with Enhanced Sensitivity Fabricated by Flexible PCB Technology For Intracranial Pressure Sensing

This paper presents a gauge pressure sensor based on the standard industrial flexible PCB (FPCB) technology. The wireless LC-resonant sensor consists of a variable sensing capacitor and a planar inductor coil fabricated in FPCB. The air gap of the sensing capacitor is extended through a pressure cha...

Full description

Saved in:
Bibliographic Details
Main Authors: Chiu, Yi, Chen, You-Zen, Hsieh, Chia-Chun, Hong, Hao-Chiao
Format: Conference Proceeding
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This paper presents a gauge pressure sensor based on the standard industrial flexible PCB (FPCB) technology. The wireless LC-resonant sensor consists of a variable sensing capacitor and a planar inductor coil fabricated in FPCB. The air gap of the sensing capacitor is extended through a pressure channel embedded in the FPCB strip to enable relative pressure measurement. Thus, the intracranial pressure (ICP) with respect to local atmosphere can be measured without other calibration means. The flat FPCB structure also facilitates the subdural implantation procedures. Experimental results showed that the proposed gauge pressure sensors with such open sensing structures had much higher frequency sensitivity (0.29%/mmHg) compared with sensors with sealed sensing cavities (0.005%/mmHg) and achieved a resolution of 0.51 mmHg.
ISSN:2168-9229
DOI:10.1109/SENSORS43011.2019.8956931