Noise Adaptive Beamforming for Linear Array Photoacoustic Imaging
The delay-and-sum (DAS) algorithm is widely used for beamforming in linear array photoacoustic (PA) imaging systems and is characterized by fast execution. However, these algorithms suffer from various drawbacks, such as low resolution, low contrast, high sidelobe artifacts, and lack of visual coher...
Saved in:
Published in: | IEEE transactions on instrumentation and measurement 2021, Vol.70, p.1-11 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The delay-and-sum (DAS) algorithm is widely used for beamforming in linear array photoacoustic (PA) imaging systems and is characterized by fast execution. However, these algorithms suffer from various drawbacks, such as low resolution, low contrast, high sidelobe artifacts, and lack of visual coherence. More recently, adaptive weighting was introduced to improve the reconstruction image quality. Unfortunately, the existing state-of-the-art adaptive beamforming algorithms are computationally expensive and do not consider the specific noise characteristics of the acquired ultrasonic signal. In this article, we present a new adaptive weighting factor named the variational coherence factor (VCF), which takes into account the noise level variations of radio frequency data. The proposed technique provides superior results in terms of image resolution, sidelobe reduction, signal-to-noise ratio (SNR), and contrast level improvement. The quantitative results of the microsphere phantom imaging show that the proposed VCF-assisted DAS method leads to 55% and 25% improvements in full-width-at-half-maximum (FWHM) and 57% and 32% improvement in SNR, respectively, compared to the state-of-the-art DAS-based methods. The results demonstrate that the proposed method can effectively improve the reconstructed image quality and deliver satisfactory imaging performance even with a limited number of sensor elements. The proposed method can potentially reduce the instrumentation cost of the PA imaging system and contribute toward the clinical translation of the modality. |
---|---|
ISSN: | 0018-9456 1557-9662 |
DOI: | 10.1109/TIM.2021.3103260 |