Loading…
Ramsey Spectroscopy in a Micro-Fabricated Rb Vapor Cell for Miniature Atomic Clocks
We present a prospective experimental study on double-resonance Ramsey spectroscopy in a micro fabricated Rb vapor cell, in view of future miniature atomic clocks with improved stability performance. Ramsey signals with a narrow central fringe width of 1.5 kHz combined with a competitive contrast of...
Saved in:
Main Authors: | , , , , , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We present a prospective experimental study on double-resonance Ramsey spectroscopy in a micro fabricated Rb vapor cell, in view of future miniature atomic clocks with improved stability performance. Ramsey signals with a narrow central fringe width of 1.5 kHz combined with a competitive contrast of > 5 % are observed, resulting in a clock short-term stability of ≈ 1×10 -11 τ −1/2 . The employed pulsed optical pumping (POP) scheme bears promise for reduced light-shift effects and thus for improved clock stability on medium- to long-term timescales. |
---|---|
ISSN: | 2327-1949 |
DOI: | 10.1109/EFTF/IFCS52194.2021.9604252 |