Loading…
Quantitative IVUS blood flow using an array catheter
In recent years, a new method to measure transverse blood flow, based on the decorrelation of radiofrequency (RF) signals has been introduced. In this paper, we studied the decorrelation characteristics of transverse blood flow using an intravascular ultrasound (IVUS) array catheter by means of comp...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In recent years, a new method to measure transverse blood flow, based on the decorrelation of radiofrequency (RF) signals has been introduced. In this paper, we studied the decorrelation characteristics of transverse blood flow using an intravascular ultrasound (IVUS) array catheter by means of computer modeling. Blood was modeled first as randomly located point scatterers and then as aggregates moving across the ultrasound beam. The spread of flow and flow gradient were then simulated. IVUS flow measurements were used to study the influence of the ring-down size on flow measurements. Coronary flow reserve (CFR) was computed from this data and from a calibrated transonic flowmeter. The decorrelation pattern for point scatterers, aggregates and nonhomogeneous flow is in agreement with theoretical predictions from sound field calculations. Underestimation of flow occurs by ring-down obscuring but good agreement between IVUS and flowmeter CFR was found. |
---|---|
ISSN: | 0276-6547 |
DOI: | 10.1109/CIC.2001.977577 |