Loading…
Accurate Calibration of Mercury Vapour Indicators for Occupational Exposure Measurements using a Dynamic Mercury Vapour Generator
An accurate and rapid calibration procedure for mercury vapour indicators used for occupational exposure measurements is described, using a dynamic mercury vapour generator traceable to the standard bell-jar calibration apparatus. The method provides greater flexibility, accuracy, and trace ability...
Saved in:
Published in: | Instrumentation science & technology 2008-11, Vol.36 (6), p.611-622 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An accurate and rapid calibration procedure for mercury vapour indicators used for occupational exposure measurements is described, using a dynamic mercury vapour generator traceable to the standard bell-jar calibration apparatus. The method provides greater flexibility, accuracy, and trace ability than the usual methods for the calibration of mercury vapour indicators used for occupational exposure measurements. The calibration procedure allows a correction to be applied to the mass concentration value displayed by the mercury vapour indicator. Measurement results from the mercury vapour indicator at the occupational exposure limit of 25 µg ·m
−3
, following calibration, have been estimated to have an expanded uncertainty of 8%. |
---|---|
ISSN: | 1073-9149 1525-6030 |
DOI: | 10.1080/10739140802448309 |