Catalog of x-ray spectra of Mo-, Rh-, and W-anode-based x-ray tubes from 10 to 50 kV
This work presents a comprehensive catalog of x-ray spectra measured from x-ray tubes with tungsten, molybdenum, and rhodium anodes generated at tube potentials between 10 and 50 kV in steps of 1 kV. They can serve as an input for dose calculations, image quality calculations, investigations of dete...
Saved in:
Published in: | Physics in medicine & biology 2021-06, Vol.66 (11), p.115013 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This work presents a comprehensive catalog of x-ray spectra measured from x-ray tubes with tungsten, molybdenum, and rhodium anodes generated at tube potentials between 10 and 50 kV in steps of 1 kV. They can serve as an input for dose calculations, image quality calculations, investigations of detector features, and validations of computational spectral models, among other things. The measurements are performed by means of a high-purity germanium detector-based spectrometer 1 m from the x-ray sources without any added filtration. The x-ray tubes are characterized by thin beryllium exit windows (0.15−4 mm); thus, for energies above 15 keV, the spectra recorded can be considered approximately unfiltered. This allows potential users of the catalog to computationally add any filter to the spectra in order to create special radiation qualities of their choice. To validate this option, a small number of spectra are recorded with filter materials in place whose purity and thickness are known with high precision. These spectra are compared to the corresponding spectra from the catalog obtained by means of computationally added filters. The two types of spectra agree extremely well. Several typical mammographic radiation qualities are selected to compare the spectra obtained from the catalog presented here with corresponding spectra obtained from other catalogs published by Boone
et al
(1997
Med. Phys.
24
1863–74) and Hernandez
et al
(2017
Med. Phys.
44
2148–60). In contrast to the work presented here, those spectra rely partly or fully on calculations. A quantitative comparison is made by means of typical x-ray quality descriptors such as the mean energy and the first and second half-value layer. The results obtained from the Boone catalog match those of the current catalog sufficiently well for the Mo- and Rh-anode-based spectra. However, significant differences up to 10 times the estimated uncertainties are found for the quality descriptors evaluated from the spectra of Hernandez
et al
and the W-anode based spectra of Boone
et al. |
---|---|