Actively-shielded ultrahigh field MRI/NMR superconducting magnet design
Active shielding technology has been widely applied to the superconducting magnetic resonance imaging (MRI) and nuclear magnetic resonance (NMR) magnets design, revealing excellent performance on the stray field control. For such a highly homogeneous field superconducting magnet design, an appropria...
Saved in:
Published in: | Superconductor science & technology 2022-01, Vol.35 (1), p.14001 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Active shielding technology has been widely applied to the superconducting magnetic resonance imaging (MRI) and nuclear magnetic resonance (NMR) magnets design, revealing excellent performance on the stray field control. For such a highly homogeneous field superconducting magnet design, an appropriate optimization strategy is essential to guarantee the magnetic field homogeneity in the central region and the expected 5 Gauss line range, especially for the ultrahigh field superconducting magnet. Based on the compensating field optimization method, an actively-shielded whole-body 14 T MRI magnet and an actively-shielded 1.3 GHz NMR magnet were presented, and detailed analyses were conducted to evaluate the feasibility of the designs. The developed magnet design method, coil pattern, wire arrangement, and stress/strain adjustment will be used to guide the corresponding project implementation. |
---|---|
ISSN: | 0953-2048 1361-6668 |