Loading…
A dual measurement and stimulating system to monitor and promote NNS
The present paper presents a dual system that can be used to measure and to generate non-nutritive sucking (NNS) signals. The NNS measurement is based on a pressure sensor that is pneumatically connected to a silicone pacifier. By its turn, the NNS stimulator is based on a miniature pump and a minia...
Saved in:
Published in: | Journal of physics. Conference series 2019-11, Vol.1379 (1), p.12034 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The present paper presents a dual system that can be used to measure and to generate non-nutritive sucking (NNS) signals. The NNS measurement is based on a pressure sensor that is pneumatically connected to a silicone pacifier. By its turn, the NNS stimulator is based on a miniature pump and a miniature electro-valve that are operated by opposite phased signals to produce the pulsed pressure signal that is pneumatically applied to the silicone pacifier. The paper highlights some specific issues related with the calibration of the pneumatic signal measuring chain. Processing and wireless transmission capabilities are provided by a low-cost solution based on a microcontroller. Calibration and experimental results are also included in the paper. |
---|---|
ISSN: | 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/1379/1/012034 |