Experimental checking results of mathematical modeling of the radiation environment sensor based on diamond detectors
Were conducted a series of experiments, the purpose of which had to verify the mathematical model of the radiation environment sensor. Theoretical values of the beta particles count rate from 90Sr - 90Y source registered by radiation environment sensor was compared with the experimental one. Theoret...
Saved in:
Published in: | Journal of physics. Conference series 2017-01, Vol.798 (1), p.12180 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Were conducted a series of experiments, the purpose of which had to verify the mathematical model of the radiation environment sensor. Theoretical values of the beta particles count rate from 90Sr - 90Y source registered by radiation environment sensor was compared with the experimental one. Theoretical (calculated) count rate of beta particles was found with using the developed mathematical model of the radiation environment sensor. Deviation of the calculated values of the beta particle count rate does not exceed 10% from the experimental. |
---|---|