Loading…

Experimental checking results of mathematical modeling of the radiation environment sensor based on diamond detectors

Were conducted a series of experiments, the purpose of which had to verify the mathematical model of the radiation environment sensor. Theoretical values of the beta particles count rate from 90Sr - 90Y source registered by radiation environment sensor was compared with the experimental one. Theoret...

Full description

Saved in:
Bibliographic Details
Published in:Journal of physics. Conference series 2017-01, Vol.798 (1), p.12180
Main Authors: Gladchenkov, E V, Ibragimov, R F, Kadilin, V V, Kolyubin, V A, Nedosekin, P G, Tyurin, E M, Zaharchenko, K V
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Were conducted a series of experiments, the purpose of which had to verify the mathematical model of the radiation environment sensor. Theoretical values of the beta particles count rate from 90Sr - 90Y source registered by radiation environment sensor was compared with the experimental one. Theoretical (calculated) count rate of beta particles was found with using the developed mathematical model of the radiation environment sensor. Deviation of the calculated values of the beta particle count rate does not exceed 10% from the experimental.
ISSN:1742-6588
1742-6596
DOI:10.1088/1742-6596/798/1/012180