Investigation of an Infrared Temperature Measurement System for Thermal Safety Verification of Plasma Skin Treatment Devices
In this paper, we developed a temperature measurement system based on an infrared temperature imagingmodule for thermal safety verification of a plasma skin treatment device (PSTD). We tested a pilot productof the low-temperature PSTD using the system, and the temperature increase of each plasma tor...
Saved in:
Published in: | Current optics and photonics 2017, 1(5), , pp.500-504 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this paper, we developed a temperature measurement system based on an infrared temperature imagingmodule for thermal safety verification of a plasma skin treatment device (PSTD). We tested a pilot productof the low-temperature PSTD using the system, and the temperature increase of each plasma torch waswell-monitored in real-time. Additionally, through the approximation of the temperature increase of theplasma torches, a certain limitation of the plasma treatment time on skin was established with theInternational Electrotechnical Commission (IEC) guideline. We determined an appropriate plasma treatmenttime (TSafe < 24 minutes) using the configured temperature measurement system. We believe that thetemperature measurement system has a potential to be employed for testing thermal safety and suitabilityof various medical devices and industrial instruments. KCI Citation Count: 1 |
---|---|
ISSN: | 2508-7266 2508-7274 |
DOI: | 10.3807/COPP.2017.1.5.500 |