Loading…
Accuracy Confirmation of Relaxation Time Mapping in Synthetic MRI
We tested the accuracy of the relaxation times T 1 and T 2 of MAGiC magnetic resonance imaging (MRI) by using a synthetic technique and comparing them with those obtained from conventional spin echo MRI. A phantom using a gadolinium contrast agent was produced and utilized to increase the accuracy o...
Saved in:
Published in: | Journal of the Korean Physical Society 2019, 74(9), , pp.892-900 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We tested the accuracy of the relaxation times
T
1 and
T
2 of MAGiC magnetic resonance imaging (MRI) by using a synthetic technique and comparing them with those obtained from conventional spin echo MRI. A phantom using a gadolinium contrast agent was produced and utilized to increase the accuracy of the measurement by minimizing the susceptibility effects that occurred in in-vivo experiments. This study showed that
T
2 mapping of MAGiC MRI has an error by comparing
T
1 and
T
2 of MAGiC MRI with those of conventional spin echo MRI, which provides the most accurate mapping value. |
---|---|
ISSN: | 0374-4884 1976-8524 |
DOI: | 10.3938/jkps.74.892 |