Loading…

The accuracy of the swallowing kinematic analysis at various movement velocities of the hyoid and epiglottis

To evaluate the accuracy of the swallowing kinematic analysis. To evaluate the accuracy at various velocities of movement, we developed an instrumental model of linear and rotational movement, representing the physiologic movement of the hyoid and epiglottis, respectively. A still image of 8 objects...

Full description

Saved in:
Bibliographic Details
Published in:Annals of rehabilitation medicine 2013, 37(3), , pp.320-327
Main Authors: Lee, Seung Hak, Oh, Byung-Mo, Chun, Seong Min, Lee, Jung Chan, Min, Yusun, Bang, Sang-Heum, Kim, Hee Chan, Han, Tai Ryoon
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:To evaluate the accuracy of the swallowing kinematic analysis. To evaluate the accuracy at various velocities of movement, we developed an instrumental model of linear and rotational movement, representing the physiologic movement of the hyoid and epiglottis, respectively. A still image of 8 objects was also used for measuring the length of the objects as a basic screening, and 18 movie files of the instrumental model, taken from videofluoroscopy with different velocities. The images and movie files were digitized and analyzed by an experienced examiner, who was blinded to the study. The Pearson correlation coefficients between the measured and instrumental reference values were over 0.99 (p
ISSN:2234-0645
2234-0653
DOI:10.5535/arm.2013.37.3.320