Head motion during cone-beam computed tomography: Analysis of frequency and influence on image quality
Image artifacts caused by patient motion cause problems in cone-beam computed tomography (CBCT) because they lead to distortion of the 3-dimensional reconstruction. This prospective study was performed to quantify patient movement during CBCT acquisition and its influence on image quality. In total,...
Saved in:
Published in: | Imaging science in dentistry 2020, 50(3), , pp.227-236 |
---|---|
Main Authors: | , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-c433t-4c24dae2626bd1348a064e8ab3f413c288e339d785e938849196acfe51ba75743 |
---|---|
cites | cdi_FETCH-LOGICAL-c433t-4c24dae2626bd1348a064e8ab3f413c288e339d785e938849196acfe51ba75743 |
container_end_page | 236 |
container_issue | 3 |
container_start_page | 227 |
container_title | Imaging science in dentistry |
container_volume | 50 |
creator | Moratin, Julius Berger, Moritz Rückschloss, Thomas Metzger, Karl Berger, Hannah Gottsauner, Maximilian Engel, Michael Hoffmann, Jürgen Freudlsperger, Christian Ristow, Oliver |
description | Image artifacts caused by patient motion cause problems in cone-beam computed tomography (CBCT) because they lead to distortion of the 3-dimensional reconstruction. This prospective study was performed to quantify patient movement during CBCT acquisition and its influence on image quality.
In total, 412 patients receiving CBCT imaging were equipped with a wireless head sensor system that detected inertial, gyroscopic, and magnetometric movements with 6 dimensions of freedom. The type and amplitude of movements during CBCT acquisition were evaluated and image quality was rated in 7 different anatomical regions of interest. For continuous variables, significance was calculated using the Student t-test. A linear regression model was applied to identify associations of the type and extent of motion with image quality scores. Kappa statistics were used to assess intra- and inter-rater agreement. Chi-square testing was used to analyze the impact of age and sex on head movement.
All CBCT images were acquired in a 10-month period. In 24% of the investigations, movement was recorded (acceleration: >0.10 [m/s
]; angular velocity: >0.018 [°/s]). In all examined regions of interest, head motion during CBCT acquisition resulted in significant impairment of image quality (
0.7).
Relevant head motions during CBCT imaging were frequently detected, leading to image quality loss and potentially impairing diagnosis and therapy planning. The presented data illustrate the need for digital correction algorithms and hardware to minimize motion artefacts in CBCT imaging. |
doi_str_mv | 10.5624/isd.2020.50.3.227 |
format | article |
fullrecord | <record><control><sourceid>proquest_nrf_k</sourceid><recordid>TN_cdi_nrf_kci_oai_kci_go_kr_ARTI_9579688</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2448405431</sourcerecordid><originalsourceid>FETCH-LOGICAL-c433t-4c24dae2626bd1348a064e8ab3f413c288e339d785e938849196acfe51ba75743</originalsourceid><addsrcrecordid>eNpVUU1v1DAQtRCIVqU_gAvyEQ5Z_Jk4HJBWFdBKlZBQOVuOPd6aJvbWTpDy7_F2ywrmMvM07z2P_BB6S8lGtkx8DMVtGGEVkQ3fMNa9QOeMcd50ipOXp5mxM3RZyi9SSzLVtfQ1OuO8AqnIOfLXYBye0hxSxG7JIe6wTRGaAcxUp2m_zODwnKa0y2Z_v37C22jGtYSCk8c-w-MC0a7YRIdD9OMBAa5mYTI7wI-LGcO8vkGvvBkLXD73C_Tz65e7q-vm9vu3m6vtbWMF53MjLBPOAGtZOzjKhTKkFaDMwL2g3DKlgPPedUpCz5USPe1bYz1IOphOdoJfoA9H35i9frBBJxOe-i7ph6y3P-5udC-7vlWqcj8fuftlmMBZiHM2o97nenlen5T_b2K4rz6_dSdJS3pZDd4_G-RUf6HMegrFwjiaCGkpmgmhBJGC00qlR6rNqZQM_vQMJfoQp65x6kOcWhLNdY2zat79e99J8Tc8_gevs5y1</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2448405431</pqid></control><display><type>article</type><title>Head motion during cone-beam computed tomography: Analysis of frequency and influence on image quality</title><source>PubMed Central</source><creator>Moratin, Julius ; Berger, Moritz ; Rückschloss, Thomas ; Metzger, Karl ; Berger, Hannah ; Gottsauner, Maximilian ; Engel, Michael ; Hoffmann, Jürgen ; Freudlsperger, Christian ; Ristow, Oliver</creator><creatorcontrib>Moratin, Julius ; Berger, Moritz ; Rückschloss, Thomas ; Metzger, Karl ; Berger, Hannah ; Gottsauner, Maximilian ; Engel, Michael ; Hoffmann, Jürgen ; Freudlsperger, Christian ; Ristow, Oliver</creatorcontrib><description>Image artifacts caused by patient motion cause problems in cone-beam computed tomography (CBCT) because they lead to distortion of the 3-dimensional reconstruction. This prospective study was performed to quantify patient movement during CBCT acquisition and its influence on image quality.
In total, 412 patients receiving CBCT imaging were equipped with a wireless head sensor system that detected inertial, gyroscopic, and magnetometric movements with 6 dimensions of freedom. The type and amplitude of movements during CBCT acquisition were evaluated and image quality was rated in 7 different anatomical regions of interest. For continuous variables, significance was calculated using the Student t-test. A linear regression model was applied to identify associations of the type and extent of motion with image quality scores. Kappa statistics were used to assess intra- and inter-rater agreement. Chi-square testing was used to analyze the impact of age and sex on head movement.
All CBCT images were acquired in a 10-month period. In 24% of the investigations, movement was recorded (acceleration: >0.10 [m/s
]; angular velocity: >0.018 [°/s]). In all examined regions of interest, head motion during CBCT acquisition resulted in significant impairment of image quality (
<0.001). Movement in the horizontal and vertical axes was most relevant for image quality (R
>0.7).
Relevant head motions during CBCT imaging were frequently detected, leading to image quality loss and potentially impairing diagnosis and therapy planning. The presented data illustrate the need for digital correction algorithms and hardware to minimize motion artefacts in CBCT imaging.</description><identifier>ISSN: 2233-7822</identifier><identifier>EISSN: 2233-7830</identifier><identifier>DOI: 10.5624/isd.2020.50.3.227</identifier><identifier>PMID: 33005580</identifier><language>eng</language><publisher>Korea (South): Korean Academy of Oral and Maxillofacial Radiology</publisher><subject>Original ; 치의학</subject><ispartof>Imaging Science in Dentistry, 2020, 50(3), , pp.227-236</ispartof><rights>Copyright © 2020 by Korean Academy of Oral and Maxillofacial Radiology.</rights><rights>Copyright © 2020 by Korean Academy of Oral and Maxillofacial Radiology 2020 Korean Academy of Oral and Maxillofacial Radiology</rights><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c433t-4c24dae2626bd1348a064e8ab3f413c288e339d785e938849196acfe51ba75743</citedby><cites>FETCH-LOGICAL-c433t-4c24dae2626bd1348a064e8ab3f413c288e339d785e938849196acfe51ba75743</cites><orcidid>0000-0002-7955-9194 ; 0000-0002-3911-262X ; 0000-0002-5187-1362 ; 0000-0002-2743-0065 ; 0000-0002-0278-7228 ; 0000-0003-4662-046X ; 0000-0003-3623-3183 ; 0000-0003-1100-6742 ; 0000-0002-4289-0457 ; 0000-0002-1322-755X</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC7506095/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC7506095/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,723,776,780,881,27903,27904,53769,53771</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33005580$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink><backlink>$$Uhttps://www.kci.go.kr/kciportal/ci/sereArticleSearch/ciSereArtiView.kci?sereArticleSearchBean.artiId=ART002629535$$DAccess content in National Research Foundation of Korea (NRF)$$Hfree_for_read</backlink></links><search><creatorcontrib>Moratin, Julius</creatorcontrib><creatorcontrib>Berger, Moritz</creatorcontrib><creatorcontrib>Rückschloss, Thomas</creatorcontrib><creatorcontrib>Metzger, Karl</creatorcontrib><creatorcontrib>Berger, Hannah</creatorcontrib><creatorcontrib>Gottsauner, Maximilian</creatorcontrib><creatorcontrib>Engel, Michael</creatorcontrib><creatorcontrib>Hoffmann, Jürgen</creatorcontrib><creatorcontrib>Freudlsperger, Christian</creatorcontrib><creatorcontrib>Ristow, Oliver</creatorcontrib><title>Head motion during cone-beam computed tomography: Analysis of frequency and influence on image quality</title><title>Imaging science in dentistry</title><addtitle>Imaging Sci Dent</addtitle><description>Image artifacts caused by patient motion cause problems in cone-beam computed tomography (CBCT) because they lead to distortion of the 3-dimensional reconstruction. This prospective study was performed to quantify patient movement during CBCT acquisition and its influence on image quality.
In total, 412 patients receiving CBCT imaging were equipped with a wireless head sensor system that detected inertial, gyroscopic, and magnetometric movements with 6 dimensions of freedom. The type and amplitude of movements during CBCT acquisition were evaluated and image quality was rated in 7 different anatomical regions of interest. For continuous variables, significance was calculated using the Student t-test. A linear regression model was applied to identify associations of the type and extent of motion with image quality scores. Kappa statistics were used to assess intra- and inter-rater agreement. Chi-square testing was used to analyze the impact of age and sex on head movement.
All CBCT images were acquired in a 10-month period. In 24% of the investigations, movement was recorded (acceleration: >0.10 [m/s
]; angular velocity: >0.018 [°/s]). In all examined regions of interest, head motion during CBCT acquisition resulted in significant impairment of image quality (
<0.001). Movement in the horizontal and vertical axes was most relevant for image quality (R
>0.7).
Relevant head motions during CBCT imaging were frequently detected, leading to image quality loss and potentially impairing diagnosis and therapy planning. The presented data illustrate the need for digital correction algorithms and hardware to minimize motion artefacts in CBCT imaging.</description><subject>Original</subject><subject>치의학</subject><issn>2233-7822</issn><issn>2233-7830</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><recordid>eNpVUU1v1DAQtRCIVqU_gAvyEQ5Z_Jk4HJBWFdBKlZBQOVuOPd6aJvbWTpDy7_F2ywrmMvM07z2P_BB6S8lGtkx8DMVtGGEVkQ3fMNa9QOeMcd50ipOXp5mxM3RZyi9SSzLVtfQ1OuO8AqnIOfLXYBye0hxSxG7JIe6wTRGaAcxUp2m_zODwnKa0y2Z_v37C22jGtYSCk8c-w-MC0a7YRIdD9OMBAa5mYTI7wI-LGcO8vkGvvBkLXD73C_Tz65e7q-vm9vu3m6vtbWMF53MjLBPOAGtZOzjKhTKkFaDMwL2g3DKlgPPedUpCz5USPe1bYz1IOphOdoJfoA9H35i9frBBJxOe-i7ph6y3P-5udC-7vlWqcj8fuftlmMBZiHM2o97nenlen5T_b2K4rz6_dSdJS3pZDd4_G-RUf6HMegrFwjiaCGkpmgmhBJGC00qlR6rNqZQM_vQMJfoQp65x6kOcWhLNdY2zat79e99J8Tc8_gevs5y1</recordid><startdate>20200901</startdate><enddate>20200901</enddate><creator>Moratin, Julius</creator><creator>Berger, Moritz</creator><creator>Rückschloss, Thomas</creator><creator>Metzger, Karl</creator><creator>Berger, Hannah</creator><creator>Gottsauner, Maximilian</creator><creator>Engel, Michael</creator><creator>Hoffmann, Jürgen</creator><creator>Freudlsperger, Christian</creator><creator>Ristow, Oliver</creator><general>Korean Academy of Oral and Maxillofacial Radiology</general><general>대한영상치의학회</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><scope>ACYCR</scope><orcidid>https://orcid.org/0000-0002-7955-9194</orcidid><orcidid>https://orcid.org/0000-0002-3911-262X</orcidid><orcidid>https://orcid.org/0000-0002-5187-1362</orcidid><orcidid>https://orcid.org/0000-0002-2743-0065</orcidid><orcidid>https://orcid.org/0000-0002-0278-7228</orcidid><orcidid>https://orcid.org/0000-0003-4662-046X</orcidid><orcidid>https://orcid.org/0000-0003-3623-3183</orcidid><orcidid>https://orcid.org/0000-0003-1100-6742</orcidid><orcidid>https://orcid.org/0000-0002-4289-0457</orcidid><orcidid>https://orcid.org/0000-0002-1322-755X</orcidid></search><sort><creationdate>20200901</creationdate><title>Head motion during cone-beam computed tomography: Analysis of frequency and influence on image quality</title><author>Moratin, Julius ; Berger, Moritz ; Rückschloss, Thomas ; Metzger, Karl ; Berger, Hannah ; Gottsauner, Maximilian ; Engel, Michael ; Hoffmann, Jürgen ; Freudlsperger, Christian ; Ristow, Oliver</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c433t-4c24dae2626bd1348a064e8ab3f413c288e339d785e938849196acfe51ba75743</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Original</topic><topic>치의학</topic><toplevel>online_resources</toplevel><creatorcontrib>Moratin, Julius</creatorcontrib><creatorcontrib>Berger, Moritz</creatorcontrib><creatorcontrib>Rückschloss, Thomas</creatorcontrib><creatorcontrib>Metzger, Karl</creatorcontrib><creatorcontrib>Berger, Hannah</creatorcontrib><creatorcontrib>Gottsauner, Maximilian</creatorcontrib><creatorcontrib>Engel, Michael</creatorcontrib><creatorcontrib>Hoffmann, Jürgen</creatorcontrib><creatorcontrib>Freudlsperger, Christian</creatorcontrib><creatorcontrib>Ristow, Oliver</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><collection>Korean Citation Index</collection><jtitle>Imaging science in dentistry</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Moratin, Julius</au><au>Berger, Moritz</au><au>Rückschloss, Thomas</au><au>Metzger, Karl</au><au>Berger, Hannah</au><au>Gottsauner, Maximilian</au><au>Engel, Michael</au><au>Hoffmann, Jürgen</au><au>Freudlsperger, Christian</au><au>Ristow, Oliver</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Head motion during cone-beam computed tomography: Analysis of frequency and influence on image quality</atitle><jtitle>Imaging science in dentistry</jtitle><addtitle>Imaging Sci Dent</addtitle><date>2020-09-01</date><risdate>2020</risdate><volume>50</volume><issue>3</issue><spage>227</spage><epage>236</epage><pages>227-236</pages><issn>2233-7822</issn><eissn>2233-7830</eissn><abstract>Image artifacts caused by patient motion cause problems in cone-beam computed tomography (CBCT) because they lead to distortion of the 3-dimensional reconstruction. This prospective study was performed to quantify patient movement during CBCT acquisition and its influence on image quality.
In total, 412 patients receiving CBCT imaging were equipped with a wireless head sensor system that detected inertial, gyroscopic, and magnetometric movements with 6 dimensions of freedom. The type and amplitude of movements during CBCT acquisition were evaluated and image quality was rated in 7 different anatomical regions of interest. For continuous variables, significance was calculated using the Student t-test. A linear regression model was applied to identify associations of the type and extent of motion with image quality scores. Kappa statistics were used to assess intra- and inter-rater agreement. Chi-square testing was used to analyze the impact of age and sex on head movement.
All CBCT images were acquired in a 10-month period. In 24% of the investigations, movement was recorded (acceleration: >0.10 [m/s
]; angular velocity: >0.018 [°/s]). In all examined regions of interest, head motion during CBCT acquisition resulted in significant impairment of image quality (
<0.001). Movement in the horizontal and vertical axes was most relevant for image quality (R
>0.7).
Relevant head motions during CBCT imaging were frequently detected, leading to image quality loss and potentially impairing diagnosis and therapy planning. The presented data illustrate the need for digital correction algorithms and hardware to minimize motion artefacts in CBCT imaging.</abstract><cop>Korea (South)</cop><pub>Korean Academy of Oral and Maxillofacial Radiology</pub><pmid>33005580</pmid><doi>10.5624/isd.2020.50.3.227</doi><tpages>10</tpages><orcidid>https://orcid.org/0000-0002-7955-9194</orcidid><orcidid>https://orcid.org/0000-0002-3911-262X</orcidid><orcidid>https://orcid.org/0000-0002-5187-1362</orcidid><orcidid>https://orcid.org/0000-0002-2743-0065</orcidid><orcidid>https://orcid.org/0000-0002-0278-7228</orcidid><orcidid>https://orcid.org/0000-0003-4662-046X</orcidid><orcidid>https://orcid.org/0000-0003-3623-3183</orcidid><orcidid>https://orcid.org/0000-0003-1100-6742</orcidid><orcidid>https://orcid.org/0000-0002-4289-0457</orcidid><orcidid>https://orcid.org/0000-0002-1322-755X</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 2233-7822 |
ispartof | Imaging Science in Dentistry, 2020, 50(3), , pp.227-236 |
issn | 2233-7822 2233-7830 |
language | eng |
recordid | cdi_nrf_kci_oai_kci_go_kr_ARTI_9579688 |
source | PubMed Central |
subjects | Original 치의학 |
title | Head motion during cone-beam computed tomography: Analysis of frequency and influence on image quality |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-27T22%3A11%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_nrf_k&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Head%20motion%20during%20cone-beam%20computed%20tomography:%20Analysis%20of%20frequency%20and%20influence%20on%20image%20quality&rft.jtitle=Imaging%20science%20in%20dentistry&rft.au=Moratin,%20Julius&rft.date=2020-09-01&rft.volume=50&rft.issue=3&rft.spage=227&rft.epage=236&rft.pages=227-236&rft.issn=2233-7822&rft.eissn=2233-7830&rft_id=info:doi/10.5624/isd.2020.50.3.227&rft_dat=%3Cproquest_nrf_k%3E2448405431%3C/proquest_nrf_k%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c433t-4c24dae2626bd1348a064e8ab3f413c288e339d785e938849196acfe51ba75743%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2448405431&rft_id=info:pmid/33005580&rfr_iscdi=true |