On-site water level measurement method based on wavelength division multiplexing for harsh environments in nuclear power plants
A simple water level measurement method based on wavelength division multiplexing (WDM) is proposed and demonstrated. The measurement principle is based on the change of Fresnel reflection occurring at the end facet of the optical fiber tip (OFT). To increase the spatial resolution of water level se...
Saved in:
Published in: | Nuclear engineering and technology 2020, 52(12), , pp.2847-2851 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A simple water level measurement method based on wavelength division multiplexing (WDM) is proposed and demonstrated. The measurement principle is based on the change of Fresnel reflection occurring at the end facet of the optical fiber tip (OFT). To increase the spatial resolution of water level sensing, a broadband light source (BLS) and an arrayed waveguide grating (AWG) are employed. The OFTs are multiplexed with the dedicated wavelength channels of AWG. By measuring all of the reflection powers reflected at the OFTs with a proposed on-site reflectometer, the water level can be monitored continuously for a fast emergency response. Moreover, it can be implemented easily with the commercially available optical components and devices with the simple configuration. |
---|---|