Analysis of the EMCCD point-source response using x-rays

Electron Multiplying Charge Coupled Devices, EMCCD are used as x-ray detectors. The NSLS-II Soft Inelastic x-ray Scattering (SIX) beam line has two EMCCDs for x-ray detection in the spectrometer arm. The spectrometer with high resolving power disperses x-rays vertically. The x-ray vertical position...

Full description

Saved in:
Bibliographic Details
Published in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2021-01, Vol.985 (C)
Main Authors: Kotov, I. V., Hall, S., Gopinath, D., Barbour, A., Li, J., Gu, Y., Holland, K., Holland, A., Jarrige, I., Pelliciari, J., Soman, M., Wilkins, S., Bisogni, V.
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page
container_issue C
container_start_page
container_title Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment
container_volume 985
creator Kotov, I. V.
Hall, S.
Gopinath, D.
Barbour, A.
Li, J.
Gu, Y.
Holland, K.
Holland, A.
Jarrige, I.
Pelliciari, J.
Soman, M.
Wilkins, S.
Bisogni, V.
description Electron Multiplying Charge Coupled Devices, EMCCD are used as x-ray detectors. The NSLS-II Soft Inelastic x-ray Scattering (SIX) beam line has two EMCCDs for x-ray detection in the spectrometer arm. The spectrometer with high resolving power disperses x-rays vertically. The x-ray vertical position on the sensor plane is related to its energy. This allows for very accurate x-ray energy measurements through x-ray coordinates. X-rays interact with silicon and create a number of electron-hole pairs proportional to the x-ray energy. Electrons drift and diffuse toward pixel gates and are collected there. The diffused electrons form a charge cloud distributed over several neighboring pixels. This charge sharing enables coordinate measurements with accuracy better than the pixel pitch. The charge distribution shape has to be taken into account to achieve ultimate accuracy in coordinate measurements. In this paper, we present a method of the charge distribution shape analysis and demonstrate its applications. The drift and diffusion of electrons ftom the point of generation to pixel gates results in the bell-shaped electron cloud usually approximated by Gaussian shape. The number of electrons collected under a pixel is proportional to the shape function integral. These electron packets get transferred to the sense node of the output amplifier. The transfer process could introduce distortions to the original charge distribution. For example, during transfers, electrons in the packet could be exposed to traps if they are present in the sensor. The trapping and later the release processes distort the apparent shape of the charge distribution. Therefore, deviations of the charge distribution shape ftom the originally symmetrical form can indicate the presence of trap centers in the sensor and can be used for sensor diagnostics.
format article
fullrecord <record><control><sourceid>osti</sourceid><recordid>TN_cdi_osti_scitechconnect_1776936</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1776936</sourcerecordid><originalsourceid>FETCH-osti_scitechconnect_17769363</originalsourceid><addsrcrecordid>eNqNyr0KwjAUQOEgCtbqO1zcA2mL-RmlVlzc3EsItzZSktKbgn17HXwAz_ItZ8WyQquSm5OSa5aJQmpuhCi3bEf0Et-M0hnT52CHhTxB7CD1CM29ri8wRh8SpzhPDmFCGmMghJl8eMKbT3ahPdt0diA8_MzZ8do86huPlHxLzid0vYshoEttoZQ0laz-mj6Hfjbl</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Analysis of the EMCCD point-source response using x-rays</title><source>ScienceDirect Freedom Collection 2022-2024</source><creator>Kotov, I. V. ; Hall, S. ; Gopinath, D. ; Barbour, A. ; Li, J. ; Gu, Y. ; Holland, K. ; Holland, A. ; Jarrige, I. ; Pelliciari, J. ; Soman, M. ; Wilkins, S. ; Bisogni, V.</creator><creatorcontrib>Kotov, I. V. ; Hall, S. ; Gopinath, D. ; Barbour, A. ; Li, J. ; Gu, Y. ; Holland, K. ; Holland, A. ; Jarrige, I. ; Pelliciari, J. ; Soman, M. ; Wilkins, S. ; Bisogni, V. ; Brookhaven National Lab. (BNL), Upton, NY (United States)</creatorcontrib><description>Electron Multiplying Charge Coupled Devices, EMCCD are used as x-ray detectors. The NSLS-II Soft Inelastic x-ray Scattering (SIX) beam line has two EMCCDs for x-ray detection in the spectrometer arm. The spectrometer with high resolving power disperses x-rays vertically. The x-ray vertical position on the sensor plane is related to its energy. This allows for very accurate x-ray energy measurements through x-ray coordinates. X-rays interact with silicon and create a number of electron-hole pairs proportional to the x-ray energy. Electrons drift and diffuse toward pixel gates and are collected there. The diffused electrons form a charge cloud distributed over several neighboring pixels. This charge sharing enables coordinate measurements with accuracy better than the pixel pitch. The charge distribution shape has to be taken into account to achieve ultimate accuracy in coordinate measurements. In this paper, we present a method of the charge distribution shape analysis and demonstrate its applications. The drift and diffusion of electrons ftom the point of generation to pixel gates results in the bell-shaped electron cloud usually approximated by Gaussian shape. The number of electrons collected under a pixel is proportional to the shape function integral. These electron packets get transferred to the sense node of the output amplifier. The transfer process could introduce distortions to the original charge distribution. For example, during transfers, electrons in the packet could be exposed to traps if they are present in the sensor. The trapping and later the release processes distort the apparent shape of the charge distribution. Therefore, deviations of the charge distribution shape ftom the originally symmetrical form can indicate the presence of trap centers in the sensor and can be used for sensor diagnostics.</description><identifier>ISSN: 0168-9002</identifier><identifier>EISSN: 1872-9576</identifier><language>eng</language><publisher>Netherlands: Elsevier</publisher><subject>Charge distribution shape ; Charge transfer efficiency ; EMCCD ; OTHER INSTRUMENTATION ; X-ray detectors</subject><ispartof>Nuclear instruments &amp; methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 2021-01, Vol.985 (C)</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><orcidid>0000000273999930 ; 0000000228185065 ; 0000000253728332 ; 0000000278232341 ; 0000000315087746 ; 0000000328919310</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885</link.rule.ids><backlink>$$Uhttps://www.osti.gov/biblio/1776936$$D View this record in Osti.gov$$Hfree_for_read</backlink></links><search><creatorcontrib>Kotov, I. V.</creatorcontrib><creatorcontrib>Hall, S.</creatorcontrib><creatorcontrib>Gopinath, D.</creatorcontrib><creatorcontrib>Barbour, A.</creatorcontrib><creatorcontrib>Li, J.</creatorcontrib><creatorcontrib>Gu, Y.</creatorcontrib><creatorcontrib>Holland, K.</creatorcontrib><creatorcontrib>Holland, A.</creatorcontrib><creatorcontrib>Jarrige, I.</creatorcontrib><creatorcontrib>Pelliciari, J.</creatorcontrib><creatorcontrib>Soman, M.</creatorcontrib><creatorcontrib>Wilkins, S.</creatorcontrib><creatorcontrib>Bisogni, V.</creatorcontrib><creatorcontrib>Brookhaven National Lab. (BNL), Upton, NY (United States)</creatorcontrib><title>Analysis of the EMCCD point-source response using x-rays</title><title>Nuclear instruments &amp; methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment</title><description>Electron Multiplying Charge Coupled Devices, EMCCD are used as x-ray detectors. The NSLS-II Soft Inelastic x-ray Scattering (SIX) beam line has two EMCCDs for x-ray detection in the spectrometer arm. The spectrometer with high resolving power disperses x-rays vertically. The x-ray vertical position on the sensor plane is related to its energy. This allows for very accurate x-ray energy measurements through x-ray coordinates. X-rays interact with silicon and create a number of electron-hole pairs proportional to the x-ray energy. Electrons drift and diffuse toward pixel gates and are collected there. The diffused electrons form a charge cloud distributed over several neighboring pixels. This charge sharing enables coordinate measurements with accuracy better than the pixel pitch. The charge distribution shape has to be taken into account to achieve ultimate accuracy in coordinate measurements. In this paper, we present a method of the charge distribution shape analysis and demonstrate its applications. The drift and diffusion of electrons ftom the point of generation to pixel gates results in the bell-shaped electron cloud usually approximated by Gaussian shape. The number of electrons collected under a pixel is proportional to the shape function integral. These electron packets get transferred to the sense node of the output amplifier. The transfer process could introduce distortions to the original charge distribution. For example, during transfers, electrons in the packet could be exposed to traps if they are present in the sensor. The trapping and later the release processes distort the apparent shape of the charge distribution. Therefore, deviations of the charge distribution shape ftom the originally symmetrical form can indicate the presence of trap centers in the sensor and can be used for sensor diagnostics.</description><subject>Charge distribution shape</subject><subject>Charge transfer efficiency</subject><subject>EMCCD</subject><subject>OTHER INSTRUMENTATION</subject><subject>X-ray detectors</subject><issn>0168-9002</issn><issn>1872-9576</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNqNyr0KwjAUQOEgCtbqO1zcA2mL-RmlVlzc3EsItzZSktKbgn17HXwAz_ItZ8WyQquSm5OSa5aJQmpuhCi3bEf0Et-M0hnT52CHhTxB7CD1CM29ri8wRh8SpzhPDmFCGmMghJl8eMKbT3ahPdt0diA8_MzZ8do86huPlHxLzid0vYshoEttoZQ0laz-mj6Hfjbl</recordid><startdate>20210101</startdate><enddate>20210101</enddate><creator>Kotov, I. V.</creator><creator>Hall, S.</creator><creator>Gopinath, D.</creator><creator>Barbour, A.</creator><creator>Li, J.</creator><creator>Gu, Y.</creator><creator>Holland, K.</creator><creator>Holland, A.</creator><creator>Jarrige, I.</creator><creator>Pelliciari, J.</creator><creator>Soman, M.</creator><creator>Wilkins, S.</creator><creator>Bisogni, V.</creator><general>Elsevier</general><scope>OTOTI</scope><orcidid>https://orcid.org/0000000273999930</orcidid><orcidid>https://orcid.org/0000000228185065</orcidid><orcidid>https://orcid.org/0000000253728332</orcidid><orcidid>https://orcid.org/0000000278232341</orcidid><orcidid>https://orcid.org/0000000315087746</orcidid><orcidid>https://orcid.org/0000000328919310</orcidid></search><sort><creationdate>20210101</creationdate><title>Analysis of the EMCCD point-source response using x-rays</title><author>Kotov, I. V. ; Hall, S. ; Gopinath, D. ; Barbour, A. ; Li, J. ; Gu, Y. ; Holland, K. ; Holland, A. ; Jarrige, I. ; Pelliciari, J. ; Soman, M. ; Wilkins, S. ; Bisogni, V.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-osti_scitechconnect_17769363</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Charge distribution shape</topic><topic>Charge transfer efficiency</topic><topic>EMCCD</topic><topic>OTHER INSTRUMENTATION</topic><topic>X-ray detectors</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Kotov, I. V.</creatorcontrib><creatorcontrib>Hall, S.</creatorcontrib><creatorcontrib>Gopinath, D.</creatorcontrib><creatorcontrib>Barbour, A.</creatorcontrib><creatorcontrib>Li, J.</creatorcontrib><creatorcontrib>Gu, Y.</creatorcontrib><creatorcontrib>Holland, K.</creatorcontrib><creatorcontrib>Holland, A.</creatorcontrib><creatorcontrib>Jarrige, I.</creatorcontrib><creatorcontrib>Pelliciari, J.</creatorcontrib><creatorcontrib>Soman, M.</creatorcontrib><creatorcontrib>Wilkins, S.</creatorcontrib><creatorcontrib>Bisogni, V.</creatorcontrib><creatorcontrib>Brookhaven National Lab. (BNL), Upton, NY (United States)</creatorcontrib><collection>OSTI.GOV</collection><jtitle>Nuclear instruments &amp; methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Kotov, I. V.</au><au>Hall, S.</au><au>Gopinath, D.</au><au>Barbour, A.</au><au>Li, J.</au><au>Gu, Y.</au><au>Holland, K.</au><au>Holland, A.</au><au>Jarrige, I.</au><au>Pelliciari, J.</au><au>Soman, M.</au><au>Wilkins, S.</au><au>Bisogni, V.</au><aucorp>Brookhaven National Lab. (BNL), Upton, NY (United States)</aucorp><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Analysis of the EMCCD point-source response using x-rays</atitle><jtitle>Nuclear instruments &amp; methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment</jtitle><date>2021-01-01</date><risdate>2021</risdate><volume>985</volume><issue>C</issue><issn>0168-9002</issn><eissn>1872-9576</eissn><abstract>Electron Multiplying Charge Coupled Devices, EMCCD are used as x-ray detectors. The NSLS-II Soft Inelastic x-ray Scattering (SIX) beam line has two EMCCDs for x-ray detection in the spectrometer arm. The spectrometer with high resolving power disperses x-rays vertically. The x-ray vertical position on the sensor plane is related to its energy. This allows for very accurate x-ray energy measurements through x-ray coordinates. X-rays interact with silicon and create a number of electron-hole pairs proportional to the x-ray energy. Electrons drift and diffuse toward pixel gates and are collected there. The diffused electrons form a charge cloud distributed over several neighboring pixels. This charge sharing enables coordinate measurements with accuracy better than the pixel pitch. The charge distribution shape has to be taken into account to achieve ultimate accuracy in coordinate measurements. In this paper, we present a method of the charge distribution shape analysis and demonstrate its applications. The drift and diffusion of electrons ftom the point of generation to pixel gates results in the bell-shaped electron cloud usually approximated by Gaussian shape. The number of electrons collected under a pixel is proportional to the shape function integral. These electron packets get transferred to the sense node of the output amplifier. The transfer process could introduce distortions to the original charge distribution. For example, during transfers, electrons in the packet could be exposed to traps if they are present in the sensor. The trapping and later the release processes distort the apparent shape of the charge distribution. Therefore, deviations of the charge distribution shape ftom the originally symmetrical form can indicate the presence of trap centers in the sensor and can be used for sensor diagnostics.</abstract><cop>Netherlands</cop><pub>Elsevier</pub><orcidid>https://orcid.org/0000000273999930</orcidid><orcidid>https://orcid.org/0000000228185065</orcidid><orcidid>https://orcid.org/0000000253728332</orcidid><orcidid>https://orcid.org/0000000278232341</orcidid><orcidid>https://orcid.org/0000000315087746</orcidid><orcidid>https://orcid.org/0000000328919310</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 0168-9002
ispartof Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 2021-01, Vol.985 (C)
issn 0168-9002
1872-9576
language eng
recordid cdi_osti_scitechconnect_1776936
source ScienceDirect Freedom Collection 2022-2024
subjects Charge distribution shape
Charge transfer efficiency
EMCCD
OTHER INSTRUMENTATION
X-ray detectors
title Analysis of the EMCCD point-source response using x-rays
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-22T20%3A02%3A13IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-osti&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Analysis%20of%20the%20EMCCD%20point-source%20response%20using%20x-rays&rft.jtitle=Nuclear%20instruments%20&%20methods%20in%20physics%20research.%20Section%20A,%20Accelerators,%20spectrometers,%20detectors%20and%20associated%20equipment&rft.au=Kotov,%20I.%20V.&rft.aucorp=Brookhaven%20National%20Lab.%20(BNL),%20Upton,%20NY%20(United%20States)&rft.date=2021-01-01&rft.volume=985&rft.issue=C&rft.issn=0168-9002&rft.eissn=1872-9576&rft_id=info:doi/&rft_dat=%3Costi%3E1776936%3C/osti%3E%3Cgrp_id%3Ecdi_FETCH-osti_scitechconnect_17769363%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true