Integration of a Multichannel Pulsed-Valve Inlet System to a Linear Quadrupole Ion Trap Mass Spectrometer for the Rapid Consecutive Introduction of Nine Reagents for Diagnostic Ion/Molecule Reactions
Gas-phase ion/molecule reactions have been used extensively for the structural elucidation of organic compounds in tandem mass spectrometry. Reagents for ion/molecule reactions can be introduced into a mass spectrometer via a continuous flow apparatus or through a pulsed inlet system. However, most...
Saved in:
Published in: | Analytical chemistry (Washington) 2019-12, Vol.91 (24), p.15652-15660 |
---|---|
Main Authors: | , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Gas-phase ion/molecule reactions have been used extensively for the structural elucidation of organic compounds in tandem mass spectrometry. Reagents for ion/molecule reactions can be introduced into a mass spectrometer via a continuous flow apparatus or through a pulsed inlet system. However, most of these approaches enable the use of only a single reagent at a time. In this work, a multichannel pulsed-valve inlet system was developed for the rapid consecutive introduction of up to nine different reagents or reagent systems into a linear quadrupole ion trap mass spectrometer for diagnostic gas-phase ion/molecule reactions. Automated triggering of the pulsed valves enabled these experiments to be performed on the high-performance liquid chromatography (HPLC) time scale. This enables high-throughput screening of several functionalities in analytes as they elute from an HPLC column. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/acs.analchem.9b03768 |