Loading…
Dilatation of stent-grafts by luminal pressures: Experimental evaluation of polytetrafluoroethylene (PTFE) and woven polyester grafts
To evaluate the effect on dilatation of three types of covering materials used for endovascular stent-grafts. Stent-grafts with three types of covering material [0.1-mm polytetrafluoroethylene (PTFE), 0.2-mm PTFE, and 0.15-mm woven polyester] were placed in a fluid circuit. For the pulsatile pressur...
Saved in:
Published in: | Cardiovascular and interventional radiology 2001-03, Vol.24 (2), p.94-98 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To evaluate the effect on dilatation of three types of covering materials used for endovascular stent-grafts.
Stent-grafts with three types of covering material [0.1-mm polytetrafluoroethylene (PTFE), 0.2-mm PTFE, and 0.15-mm woven polyester] were placed in a fluid circuit. For the pulsatile pressure test, the luminal pressure of 190/130 mmHg was loaded up to 300,000 pulses. For the static pressure test, the luminal pressure was increased from 50 to 300 mmHg at 50-mmHg increments. The percent of dilatation of each stent-graft was compared.
The 0.1-mm PTFE stent-graft was significantly more dilated than the 0.2-mm PTFE and the 0.15-mm woven polyester stent-graft (p < 0.005) in both examinations. There was no significant difference between the 0.2-mm PTFE and the 0.15-mm woven polyester stent-grafts. The dilatation of the 0.1-mm PTFE stent-graft was irreversible.
The 0.15-mm woven polyester and the 0.2-mm PTFE stent-grafts may be preferable to the 0.1-mm PTFE stent-graft with regard to dilatation and deformity. |
---|---|
ISSN: | 0174-1551 1432-086X |
DOI: | 10.1007/s002700000388 |