Automated laboratory based X-ray beamline with multi-capillary sample chamber

An automated laboratory based X-ray beamline with a multi-capillary sample chamber capable of undertaking small angle X-ray scattering measurements on a maximum of 104 samples at a time as a function of temperature between 5 and 85 °C has been developed. The modular format of the system enables the...

Full description

Saved in:
Bibliographic Details
Published in:Review of scientific instruments 2013-08, Vol.84 (8), p.085104
Main Authors: Purushothaman, S, Gauthé, B L L E, Brooks, N J, Templer, R H, Ces, O
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An automated laboratory based X-ray beamline with a multi-capillary sample chamber capable of undertaking small angle X-ray scattering measurements on a maximum of 104 samples at a time as a function of temperature between 5 and 85 °C has been developed. The modular format of the system enables the user to simultaneously equilibrate samples at eight different temperatures with an accuracy of ±0.005 °C. This system couples a rotating anode generator and 2D optoelectronic detector with Franks X-ray optics, leading to typical exposure times of less than 5 min for lyotropic liquid crystalline samples. Beamline control including sample exchange and data acquisition has been fully automated via a custom designed LabVIEW framework.
ISSN:0034-6748
1089-7623