Loading…
Photodissociation/gas diffusion/ion chromatography system for determination of total and labile cyanide in waters
An automated system for determination of total and labile cyanide in water samples has been developed. The stable metal-cyanide complexes such as Fe(CN){sub 6}{sup 3{minus}} are photodissociated in an acidic medium with an on-line pyrex glass reaction coil irradiated by an intense Hg lamp. The relea...
Saved in:
Published in: | Analytical chemistry (Washington) 1990-04, Vol.62 (7), p.766-770 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An automated system for determination of total and labile cyanide in water samples has been developed. The stable metal-cyanide complexes such as Fe(CN){sub 6}{sup 3{minus}} are photodissociated in an acidic medium with an on-line pyrex glass reaction coil irradiated by an intense Hg lamp. The released cyanide (HCN) is separated from most interferences in the sample matrix and is collected in a dilute NaOH solution by gas diffusion using a hydrophobic porous membrane separator. The cyanide ion is then separated from remaining interferences such as sulfide by ion exchange chromatography and is detected by an amperometric detector. The characteristics of the automated system were studied with solutions of free cyanide and metal-cyanide complexes. The results of cyanide determination for a number of wastewater samples obtained with the new method were compared with those obtained with the standard method. The sample throughput of the system is eight samples per hour and the detection limit for total cyanide is 0.1 {mu}g/L. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/ac00206a024 |