Helium discharge detector for quantitation of volatile organohalogen compounds

A helium discharge has been evaluated as an element-selective, gas chromatographic detector for volatile organohalogen compounds. Absolute limits of detection ranged from 3 pg for chloroform to 29 pg for p-dibromobenzene with a linear response of 10(3)-10(4). The relative response of the detector wa...

Full description

Saved in:
Bibliographic Details
Published in:Analytical chemistry (Washington) 1990-04, Vol.62 (8), p.853-857
Main Authors: Ryan, David A, Argentine, Suzanne M, Rice, Gary W
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A helium discharge has been evaluated as an element-selective, gas chromatographic detector for volatile organohalogen compounds. Absolute limits of detection ranged from 3 pg for chloroform to 29 pg for p-dibromobenzene with a linear response of 10(3)-10(4). The relative response of the detector was determined to be based solely on the mass of the halogen species present. This feature allowed for relatively simple quantitation of chlorinated and brominated haloform species in water samples by the addition of a single internal standard for calibration of the halogen response.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac00207a017