“Age” determination of irradiated materials utilizing inductively coupled plasma mass spectrometric (ICP-MS) detection

A gas pressurized extraction chromatography (GPEC) system has been developed to perform elemental separations on radioactive samples to determine total and isotopic compositions of Cs and Ba from an irradiated salt sample, fuel sample and two sealed radiation sources. The GPEC system employs compres...

Full description

Saved in:
Bibliographic Details
Published in:Journal of radioanalytical and nuclear chemistry 2009-11, Vol.282 (2), p.591-595
Main Authors: Sommers, James, Cummings, Daniel, Giglio, Jeffrey, Carney, Kevin
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A gas pressurized extraction chromatography (GPEC) system has been developed to perform elemental separations on radioactive samples to determine total and isotopic compositions of Cs and Ba from an irradiated salt sample, fuel sample and two sealed radiation sources. The GPEC system employs compressed nitrogen to move liquid through the system, compared to gravity or pumped liquids that are typically used for separations. A commercially available Sr-Resin™ was used to perform the separation for the above mentioned analytes. A 1% acetic acid solution was determined to be the best extractant for Ba. A flow rate of 0.1 mL/min was determined to be optimal for the separation of Ba. Complete recovery of the Cs and Ba was achieved, within the systematic uncertainties of the experiments.
ISSN:0236-5731
1588-2780
DOI:10.1007/s10967-009-0210-8