Biomagnetic measurements with HTS SQUID magnetometers in moderate shielded environments

An HTS SQUID magnetometer for biomagnetic measurements was implemented in a moderate noise shielded environment. The intrinsic noises of the SQUID magnetometer shielded with a tri-layer magnetic shielding cylinder were 50 fT/Hz/sup 1/2/ above 10 Hz and 100 fT/Hz/sup 1/2/ at 0.1 Hz. The moderate shie...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on applied superconductivity 2003-06, Vol.13 (2), p.381-384
Main Authors: Herng-Er Horng, Shou-Yen Hung, Jen-Tzong Jeng, Shu-Hsien Liao, Hsu, S.-C., Ji-Chuu Hwang, Hong-Chang Yang
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An HTS SQUID magnetometer for biomagnetic measurements was implemented in a moderate noise shielded environment. The intrinsic noises of the SQUID magnetometer shielded with a tri-layer magnetic shielding cylinder were 50 fT/Hz/sup 1/2/ above 10 Hz and 100 fT/Hz/sup 1/2/ at 0.1 Hz. The moderate shielded environments are constructed of mu-metal layers. With the moderate shielding the magnitude of the QRS complex of a rat specimen was found to be around 10 pT and the signal-to-noise ratio was larger than 10. The requirements for improving biomagnetic measurements in the moderate magnetic shielding are discussed.