A high precision manometric system for absolute calibrations of CO2 in dry air
A high‐precision manometric system has been developed for absolute calibrations of CO2‐in‐air mixture gas. This report describes the principle of the calibration method and evaluates the performance of the manometric apparatus. The test results show that the reproducibility of the manometric system...
Saved in:
Published in: | Journal of Geophysical Research: Atmospheres 1997-03, Vol.102 (D5), p.5885-5894 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A high‐precision manometric system has been developed for absolute calibrations of CO2‐in‐air mixture gas. This report describes the principle of the calibration method and evaluates the performance of the manometric apparatus. The test results show that the reproducibility of the manometric system for calibrating CO2‐in‐air gas mixtures is within ±0.1 μmol mol−1 in the atmospheric CO2 concentration range of 300 to 400 μmol mol−1. Preliminary measurements indicate that the agreement with the absolute World Meteorological Organization mole fraction scale, widely used for atmospheric CO2 monitoring, is also within ±0.1 μmol mol−1 in that same range. |
---|---|
ISSN: | 0148-0227 2156-2202 |
DOI: | 10.1029/96JD03764 |