Loading…
A Technique for Extending the Dynamic Range of the Dual Six-Port Network Analyzer
The dynamic range of the six-port type of automatic network analyzer is typically limited to measuring two-port devices with a transmission coefficient S/sub 12/ in the range of 0 to -60 dB. The following describes a subcarrier approach for extending the dynamic range of the dual six-port network an...
Saved in:
Published in: | IEEE transactions on microwave theory and techniques 1985-06, Vol.33 (6), p.453-459 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The dynamic range of the six-port type of automatic network analyzer is typically limited to measuring two-port devices with a transmission coefficient S/sub 12/ in the range of 0 to -60 dB. The following describes a subcarrier approach for extending the dynamic range of the dual six-port network analyzer. The subcarrier is generated by inserting a 10-kHz, biphase modulator ahead of one of the six-ports. With the subcarrier approach, measurements of S/sub 12/ in the range of -60 to -100 dB can be made. Test results are presented showing measurements of S/sub 12/ =-80 db with a precision of +-0.05 dB or better, and an accuracy of +-0.16 dB or better at 3 GHz. Measurement results are also presented showing the dynamic range achievable with thermistor and barretter detectors. Key Words: barretter power detectors; diode power detectors; impedance measurements; microwave network analyze; six-port network analyze; thermistor power detectors. |
---|---|
ISSN: | 0018-9480 1557-9670 |
DOI: | 10.1109/TMTT.1985.1133099 |