Loading…

Comparative analysis of postural control and vertical jump performance between three different measurement devices

The aim of this study was to examine the concurrent validity of the HUMAC Balance System (HBS) and Balance Trainer BTG4 (BTG) in comparison to a laboratory-grade force platform (FP) for postural control (PC) and vertical jump performance (VJP) assessment. In addition, reliability of the three device...

Full description

Saved in:
Bibliographic Details
Published in:PloS one 2019-09, Vol.14 (9), p.e0222502-e0222502
Main Authors: Blosch, Christopher, Schäfer, Robin, de Marées, Markus, Platen, Petra
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The aim of this study was to examine the concurrent validity of the HUMAC Balance System (HBS) and Balance Trainer BTG4 (BTG) in comparison to a laboratory-grade force platform (FP) for postural control (PC) and vertical jump performance (VJP) assessment. In addition, reliability of the three devices was measured for PC. Overall 22 participants (age = 37.8 ± 13.3 years; gender = 9 male, 13 female; height = 174.1 ± 10.5 cm; body mass = 75.3 ± 17.6 kg) were recruited to participate. Double and single leg standing balance tests with eyes open or closed and counter movement jumps (CMJ) were performed on two separate occasions. Reliability and concurrent validity for COP parameters and VJP were examined using intraclass correlation coefficients (ICC), Bland-Altman plots (BAP), standard error of measurement (SEM) and minimum detectable change (MDC). COP path length test-retest reliability was predominantly good to excellent for all three devices (ICC = 0.80-0.95). SEM and MDC values were high for all plates (SEM% = 8.0-15.2; MDC% = 22.8-44.5), with the HBS MDC values higher than the KIS and BTG in three of the four trials. ICC scores for concurrent validity were good to excellent for the BTG (ICC = 0.76-0.93) and moderate to good for the HBS (0.49-0.83). Band-Altman plots revealed a systematic bias for the HBS towards higher COP path length values under all conditions and for the BTG in two out of four trials towards lower values. Validity of VJP was excellent for the BTG (ICC = 1.0) and poor for the HBS (0.34), with a systematic bias towards lower values. The comparative analysis of PC and VJP revealed reliable and valid results for the BTG in comparison to a laboratory-grade force plate. The HBS showed reliable results for PC assessment with restrictions regarding its validity. Results of VJP showed that the HBS revealed deficits in the assessment of activities that require rapid, high force movements such as jumping and running. Due to the variable results of all three devices, it is recommended not to use them interchangeably.
ISSN:1932-6203
1932-6203
DOI:10.1371/journal.pone.0222502