Loading…
Fluorescence Immunoassay Based on Long Time Correlations of Number Fluctuations
We report the development of a fluorescence-based immunoassay technique relying on the physical phenomena of random number fluctuations and diffusion, which we review. By determining the autocorrelation of the fluctuations in the fluorescent intensity, this method is able to measure the amount of la...
Saved in:
Published in: | Proceedings of the National Academy of Sciences - PNAS 1980-08, Vol.77 (8), p.4904-4908 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We report the development of a fluorescence-based immunoassay technique relying on the physical phenomena of random number fluctuations and diffusion, which we review. By determining the autocorrelation of the fluctuations in the fluorescent intensity, this method is able to measure the amount of labeled antigen or antibody that is bound to micrometer-sized carrier particles in solution. The principal advantage of this technique is its insensitivity to small, fast-diffusing sources. It also discriminates against weakly fluorescent contaminants of size comparable to the carrier particles. We demonstrate these attributes by using two model systems: a human IgG assay and an idealized system consisting of polystyrene fluorescent spheres and rhodamine dye. |
---|---|
ISSN: | 0027-8424 1091-6490 |