Loading…
Hidden dangers in a simulation centre
Abstract In the article, we emphasize the need for all equipment in a simulation centre to be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve wen...
Saved in:
Published in: | Zdravniški vestnik (Ljubljana, Slovenia : 1992) Slovenia : 1992), 2011-06, Vol.80 (6) |
---|---|
Main Authors: | , |
Format: | Article |
Language: | Slovenian |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract In the article, we emphasize the need for all equipment in a simulation centre to be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve went unnoticed, resulting in slow, imperceptible leakage of gas from the cylinder. A careful soap test revealed that gas was escaping through a small nozzle on the back of the pressure regulator. The leakage was inaudible but could be felt by placing a finger on the valve nozzle. |
---|---|